Three-Dimensional Mechanical Characterization of Normal, Subclinical, and Keratoconic Corneas Using Motion Tracking Brillouin Microscopy.

Randleman J Bradley, Zhang Hongyuan, Hammoud Bassel, Dutra Barbara A L, Susanna Bianca N, Scarcelli Giuliano

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To characterize the 3-dimensional (3-D) mechanical differences between normal corneas and those with subclinical keratoconus (SKC) or clinical keratoconus (KC) using motion-tacking (MT) Brillouin microscopy.

DESIGN

Prospective cross-sectional study.

METHODS

There were 50 eyes from 50 patients evaluated, including 20 Control eyes from 20 patients, 20 SKC eyes from 20 patients, and 10 stage I or II KC eyes from 10 patients. All patients had Scheimpflug imaging for group classification and MT Brillouin microscopy using a custom-built device for mechanical evaluation. Mean and minimum MT Brillouin shift values were generated in the anterior, middle, and posterior stromal regions. Receiver operating characteristic (ROC) curves were generated for MT Brillouin metrics to determine the area under the ROC curve (AUROC), sensitivity, and specificity for each variable at each location.

RESULTS

There were no significant differences between groups for age, sex, or mean keratometry. There were significant differences between control, SKC, and KC groups for all MT Brillouin metrics, with greatest differences in anterior min (Δ 43-57 MHz, P < .001) and Mean (Δ 34-37 MHz, P < .001), middle min (Δ 44-69 MHz, P < .001) and Mean (Δ 39-54 MHz, P < .001), and least differences in posterior min (Δ 15-22 MHz, P < .001) and Mean (Δ 20-32 MHz, P < .001). Anterior min values fully separated both SKC and KC groups from controls (AUC = 1.0 for each), while anterior mean (AUC = 0.98, 1.0), middle mean (AUC = 0.96, 1.0), and middle min (AUC = 0.94, 1.0) performed better than posterior mean (AUC = 0.81, 0.95) or posterior min (AUC = 0.78, 0.88).

CONCLUSIONS

MT Brillouin microscopy identified 3-dimensional differences between controls eyes and those with subclinical or early-stage keratoconus. Differences were most pronounced in the anterior stroma and least pronounced in the posterior stroma, indicating relevant axial mechanical differences in eyes with SKC. NOTE: Publication of this article is sponsored by the American Ophthalmological Society.

Для цитирования: Randleman J Bradley, Zhang Hongyuan, Hammoud Bassel, Dutra Barbara A L, Susanna Bianca N, Scarcelli Giuliano. Three-Dimensional Mechanical Characterization of Normal, Subclinical, and Keratoconic Corneas Using Motion Tracking Brillouin Microscopy.. American journal of ophthalmology. 2025. DOI: 10.1016/j.ajo.2025.06.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.