Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols: 25-3D Scan and 50-Cornea Fine.

Romano Davide, Coco Giulia, Borgia Alfredo, Calza Stefano, Kaye Stephen, Gadhvi Kunal, Semeraro Francesco, Romano Vito

СтатьяПеревод пока не подготовлен

Аннотация

Background

This study aims to evaluate the repeatability of the Pentacam HR, comparing two different measurement modes (50-cornea fine and 25-3D scan) in patients affected by keratoconus.

Methods

Multicenter retrospective study, conducted at Eye Clinic of the ASST-Spedali Civili-University of Brescia, Italy, and St. Paul's Eye Unit, Royal Liverpool University Hospital, United Kingdom. A total of 72 eyes from 72 patients with keratoconus underwent six consecutive measurements, three using the 25-3D scan mode and three with the 50-Cornea fine mode. Measurements were made by one single observer, using the Scheimpflug corneal tomographer camera (Pentacam HR, Oculus, Wetzlar, Germany). Repeatability was assessed using the within-subject SD (Sw) statistic from a two-way analysis of variance.

Results

Both measurement modes had excellent repeatability. The interclass coefficient correlation (ICC) was excellent (>0.9) in all the parameters evaluated, apart from anterior and posterior astigmatic axes and posterior astigmatism (ICC > 0.8) and index of height asymmetry (IHA) (ICC < 0.6). However, in 18 of 29 parameters, the ICC was higher in case of 25-3D scan. Repeatability limit for Kmax was 1.00D in 25-3D scan mode and 1.02D in case of 50-cornea fine.

Conclusions

25-3D scan may be preferable to 50-Cornea fine, in view of having slightly higher ICC in case of patients with keratoconus. Repeatability limits reported may be helpful in clinical practice for assessing the progression of keratoconus.

Для цитирования: Romano Davide, Coco Giulia, Borgia Alfredo, Calza Stefano, Kaye Stephen, Gadhvi Kunal, Semeraro Francesco, Romano Vito. Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols: 25-3D Scan and 50-Cornea Fine.. Journal of clinical medicine. 2025. DOI: 10.3390/jcm14020439.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.