Malpositioned spinal instrument as a possible cause of superficial siderosis.

Aoyama T, Hida K, Yano S, Kawabori M, Iwasaki M, Iwasaki Y

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

STUDY

DESIGN

Case report.

OBJECTIVE

To report a patient with superficial siderosis as a complication after posterior fixation surgery for odontoid fracture.

SETTING

Department of Neurosurgery, Hokkaido University, Japan.

METHODS

A 36-year-old man had undergone C1-C2 posterior fixation using lamina hooks for an odontoid fracture in 1997. In 2003, he presented with hearing loss and ataxia; and in 2006, a diagnosis of superficial siderosis was made and spinal instrument malpositioning was detected.

RESULTS

The malpositioned instrument, suspected as the cause of superficial siderosis, was removed.

CONCLUSIONS

Superficial siderosis of the central nervous system is rare; it results in progressive hearing loss, cerebellar ataxia and pyramidal sign. Chronic hemorrhage in the subarachnoid space precipitates hemosiderin around the cerebellum and brainstem resulting in neurological symptoms. Recurrent hemorrhage and cervical root pathology, for example, root avulsion, are factors; the symptoms worsen gradually and result in hemostasis. Superficial siderosis because of complications from spinal instrumentation surgery is extremely rare. If the instrument is malpositioned in the subarachnoid space, we suggest its removal.

Для цитирования: Aoyama T, Hida K, Yano S, Kawabori M, Iwasaki M, Iwasaki Y. Malpositioned spinal instrument as a possible cause of superficial siderosis.. Spinal cord. 2009. DOI: 10.1038/sc.2009.35.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.