CT imaging before transcatheter aortic valve implantation (TAVI) using variable helical pitch scanning and its diagnostic performance for coronary artery disease.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Matsumoto Shunsuke, Yamada Yoshitake, Hashimoto Masahiro, Okamura Teppei, Yamada Minoru, Yashima Fumiaki, Hayashida Kentaro, Fukuda Keiichi, Jinzaki Masahiro
Аннотация
OBJECTIVES: To evaluate the effectiveness of CT before TAVI using variable helical pitch (VHP) scanning and its diagnostic performance for coronary artery disease (CAD).
METHODS: Sixty patients (84.4 ± 4.6 years) scheduled for TAVI underwent CT using VHP scanning with the contrast material (CM) volume calculated as scanning time × weight [kg] × 0.06 mL. Retrospective electrocardiography (ECG)-gated scanning was utilized to examine the thorax, and non-ECG-gated scanning of the abdomen immediately followed. We analyzed CT attenuation values of the coronary arteries, aorta, iliac and femoral arteries. The coronary CT angiography images were evaluated for the presence of stenosis (≥50 %); invasive coronary angiography served as a reference standard.
RESULTS: The average attenuations of all of the arteries were greater than 400 HU. We could evaluate the peripheral access vessels and dimensions of the ascending aorta, aortic root, and aortic annulus in all patients. The average volume of CM was 38.7 ± 8.5 mL. On per-patient and vessel analysis, CT showed 91.7 % and 89.5 % sensitivity, and 91.3 % and 97.4 % negative predictive value (NPV).
CONCLUSIONS: CT using VHP scanning with an average CM volume of 38.7 mL is useful before TAVI and had a high sensitivity and NPV in excluding obstructive CAD.
KEY POINTS: • TAVI-planning CT using variable helical pitch (VHP) scanning is useful. • The average volume of contrast material was 38.7 ± 8.5 mL. • The average attenuations of all the arteries were >400 HU. • This CT had a high sensitivity and NPV for excluding obstructive CAD.
METHODS: Sixty patients (84.4 ± 4.6 years) scheduled for TAVI underwent CT using VHP scanning with the contrast material (CM) volume calculated as scanning time × weight [kg] × 0.06 mL. Retrospective electrocardiography (ECG)-gated scanning was utilized to examine the thorax, and non-ECG-gated scanning of the abdomen immediately followed. We analyzed CT attenuation values of the coronary arteries, aorta, iliac and femoral arteries. The coronary CT angiography images were evaluated for the presence of stenosis (≥50 %); invasive coronary angiography served as a reference standard.
RESULTS: The average attenuations of all of the arteries were greater than 400 HU. We could evaluate the peripheral access vessels and dimensions of the ascending aorta, aortic root, and aortic annulus in all patients. The average volume of CM was 38.7 ± 8.5 mL. On per-patient and vessel analysis, CT showed 91.7 % and 89.5 % sensitivity, and 91.3 % and 97.4 % negative predictive value (NPV).
CONCLUSIONS: CT using VHP scanning with an average CM volume of 38.7 mL is useful before TAVI and had a high sensitivity and NPV in excluding obstructive CAD.
KEY POINTS: • TAVI-planning CT using variable helical pitch (VHP) scanning is useful. • The average volume of contrast material was 38.7 ± 8.5 mL. • The average attenuations of all the arteries were >400 HU. • This CT had a high sensitivity and NPV for excluding obstructive CAD.
Предметные рубрики MeSH
AgedAged, 80 and overAortaAortic ValveAortic Valve StenosisComputed Tomography AngiographyContrast MediaCoronary AngiographyCoronary Artery DiseaseCoronary StenosisElectrocardiographyFemaleFemoral ArteryHumansIliac ArteryMalePreoperative CareRetrospective StudiesTomography, Spiral ComputedTomography, X-Ray ComputedTranscatheter Aortic Valve Replacement