Kansas journal of medicine · 2021
Post-Biopsy Pneumothorax Incidence in Patients Treated with Biosentry™ Plug Device.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Abouodah Hythem, Werner Gregg, Fahrbach Thomas M, Fox Cameron, Mazurek Jared, Lott Jeremy, Walter Carissa N, Clark Lauren, Custer Brandon
Аннотация
INTRODUCTION: This study aimed to determine if the Biosentry™ Plug Device (BPD), a prophylactic sealant used to prevent pneumothorax after lung biopsies, reduced post-lung biopsy pneumothorax rates, and other complications compared to no device utilization.
METHODS: This single institution, retrospective cohort study included patients who received a lung biopsy in the Department of Interventional Radiology from May 1, 2015 to August 31, 2017. Data such as sex, race, ethnicity, chronic obstructive pulmonary disease status, degree of lung bullae if present, smoking status, and use of BPD were recorded. Decisions to use BPD were based on operator preference. A chi squared analysis was used with a p value greater than 0.05 considered significant.
RESULTS: The study included 521 patients who underwent a lung biopsy during the study timeframe. Of these, 74 (14.2%) received the BPD, while 447 (85.8%) did not. One-hundred ninety (36.4%) had a pneumothorax within one month of the lung biopsy. Of the total 190 that experienced pneumothorax, 36.7% of non-BPD biopsies resulted in pneumothorax, while 35.1% of BPD biopsies resulted in pneumothorax (p value = 0.7970; degrees of freedom = 1).
CONCLUSIONS: These findings indicated that BPD may not reduce pneumothorax incidence nor limit the severity of complications in patients.
METHODS: This single institution, retrospective cohort study included patients who received a lung biopsy in the Department of Interventional Radiology from May 1, 2015 to August 31, 2017. Data such as sex, race, ethnicity, chronic obstructive pulmonary disease status, degree of lung bullae if present, smoking status, and use of BPD were recorded. Decisions to use BPD were based on operator preference. A chi squared analysis was used with a p value greater than 0.05 considered significant.
RESULTS: The study included 521 patients who underwent a lung biopsy during the study timeframe. Of these, 74 (14.2%) received the BPD, while 447 (85.8%) did not. One-hundred ninety (36.4%) had a pneumothorax within one month of the lung biopsy. Of the total 190 that experienced pneumothorax, 36.7% of non-BPD biopsies resulted in pneumothorax, while 35.1% of BPD biopsies resulted in pneumothorax (p value = 0.7970; degrees of freedom = 1).