MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1167/tvst.12.3.1

Медицинская библиотека · Translational vision science & technology · 2023

The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision.

Denkinger Sylvie, Antoniou Maria-Paraskevi, Tarello Demetrio, Levi Dennis M, Backus Benjamin T, Bavelier Daphné, Chopin Adrien

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To describe two new stereoacuity tests: the eRDS v6 stereotest, a global dynamic random dot stereogram (dRDS) test, and the Vivid Vision Stereo Test version 2 (VV), a local or "contour" stereotest for virtual reality (VR) headsets; and to evaluate the tests' reliability, validity compared to a dRDS standard, and learning effects.

METHODS

Sixty-four subjects passed a battery of stereotests, including perceiving depth from RDS. Validity was evaluated relative to a tablet-based dRDS reference test, ASTEROID. Reliability and learning effects were assessed over six sessions.

RESULTS

eRDS v6 was effective at measuring small thresholds (<10 arcsec) and had a moderate correlation (0.48) with ASTEROID. Across the six sessions, test-retest reliability was good, varying from 0.84 to 0.91, but learning occurred across the first three sessions. VV did not measure stereoacuities below 15 arcsec. It had a weak correlation with ASTEROID (0.27), and test-retest reliability was poor to moderate, varying from 0.35 to 0.74; however, no learning occurred between sessions.

CONCLUSIONS

eRDS v6 is precise and reliable but shows learning effects. If repeated three times at baseline, this test is well suited as an outcome measure for testing interventions. VV is less precise, but it is easy and rapid and shows no learning. It may be useful for testing interventions in patients who have no global stereopsis.

TRANSLATIONAL RELEVANCE: eRDS v6 is well suited as an outcome measure to evaluate treatments that improve adult stereodepth perception. VV can be considered for screening patient with compromised stereovision.

Для цитирования: Denkinger Sylvie, Antoniou Maria-Paraskevi, Tarello Demetrio, Levi Dennis M, Backus Benjamin T, Bavelier Daphné, Chopin Adrien. The eRDS v6 Stereotest and the Vivid Vision Stereo Test: Two New Tests of Stereoscopic Vision.. Translational vision science & technology. 2023. DOI: 10.1167/tvst.12.3.1.