[Successful post-bypass extracorporeal circulatory assistance with the centrifugal pump].
Аннотация
Patients with reduced left ventricular function do have an increased risk of inability to be weaned off bypass after open heart surgery despite maximal pharmacologic support and intraaortic balloon counter-pulsation. Centrifugal pumps used for extracorporeal circulatory assist can maintain a patient in low cardiac output up to days without anticoagulation. We used a centrifugal pump in 3 patients: as a left ventricular assist device (LVAD) in 2 patients and right ventricular assist device (RVAD) in 1 patient. One LVAD-patient became a long-term survivor after 20 h of assist, another was bridged successfully to an open heart procedure for 2 h after papillary muscle rupture. One RVAD patient died on the operating table due to massive tracheal bleeding probably caused by pulmonary hypertension.