Midterm experience with the Jarvik 2000 axial flow left ventricular assist device.

Haj-Yahia Saleem, Birks Emma J, Rogers Paula, Bowles Christopher, Hipkins Mandy, George Robert, Amrani Mohammed, Petrou Mario, Pepper John, Dreyfus Gilles, Khaghani Asghar

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Rotary axial flow pumps have several potential advantages and disadvantages over pulsatile pumps. The Jarvik 2000 is distinctive in being intracardiac. We report our experience in 22 patients.

METHODS

The Jarvik 2000 was implanted in 15 men and 7 women. Mean age was 38.8 (range 23-59) years, preoperative diagnosis was dilated cardiomyopathy in 16, postpartum cardiomyopathy in 3, ischemic heart disease in 2, and chronic allograft failure in 1. Twenty-one patients were in New York Heart Association class IV, and 1 patient was in class III. Nineteen patients were on inotropic support, 6 were supported with an intra-aortic balloon pump, and 2 patients had been salvaged with a Centrimag (Levitronix) ventricular assist device. The median pulmonary vascular resistance was 3 Wood units; median pulmonary capillary wedge pressure was 26.6 mm Hg; and mean Cardiac Index was 1.5 L/min/m2.

RESULTS

There were 2 early deaths and 6 late deaths. The average postoperative ventilation time and Intensive Treatment Unit stay was 2.2 and 10 days, respectively. One patient required a right ventricular assist device for short-term support and another for medium-term support. Seven patients were bridged to transplant, 3 had myocardial recovery, and 4 are ongoing. Mean and total duration of support was 280.5 and 6172 days, respectively. Driveline failures were noted in 3, but there were no pump infections or failure.

CONCLUSION

The Jarvik 2000 provides satisfactory intermediate-term results as a bridge to transplant or recovery. It appears to be associated with a low rate of serious driveline or pump infections and technical failure. However, bleeding complications due to the required anticoagulation treatment frequently occurred.

Для цитирования: Haj-Yahia Saleem, Birks Emma J, Rogers Paula, Bowles Christopher, Hipkins Mandy, George Robert, Amrani Mohammed, Petrou Mario, Pepper John, Dreyfus Gilles, Khaghani Asghar. Midterm experience with the Jarvik 2000 axial flow left ventricular assist device.. The Journal of thoracic and cardiovascular surgery. 2007. DOI: 10.1016/j.jtcvs.2007.01.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.