MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000001589

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2017

Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation.

Choi June, Payne Matthew R, Campbell Luke J, Bester Christo W, Newbold Carrie, Eastwood Hayden, O'Leary Stephen J

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES/HYPOTHESIS: Cochlear implant surgery now aims to preserve residual low frequency hearing. The current research explores whether fluctuations in the electrical impedance of cochlear implant electrodes may act as a biomarker for pathological changes that lead to the delayed loss of residual hearing.

STUDY

DESIGN

Secondary analysis of a double-blinded randomized trial, where methylprednisolone was administered intravenously before cochlear implantation with a view to preserving residual hearing.

METHODS

Seventy-four patients with residual hearing after cochlear implant surgery were investigated for an impedance "spike," defined as a median rise of ≥4 kΩ across all electrodes from the baseline measurements. Spikes were related to objective and subjective hearing loss, dizziness, and tinnitus.

RESULTS

An impedance spike occurred in 14% (10/74) of enrolled patients. Three months after surgery, five patients exhibited spikes and three of these patients had a total loss of their residual hearing. 4.3% of the 69 patients without spikes lost residual hearing. At 1 year, 9 of 10 patients who exhibited spikes had lost all their residual hearing. 8.1% of the 37 patients who did not experience a spike lost their residual hearing. Seventy percent of patients exhibiting a spike also experienced vertigo. The administration of steroids at the time of surgery did not influence the occurrence of spikes.

CONCLUSION

Our results suggest that there is a relationship between a spike and the loss of residual hearing. It seems that rises in impedance can reflect pathology within the inner ear and predict the future loss of residual hearing.

Для цитирования: Choi June, Payne Matthew R, Campbell Luke J, Bester Christo W, Newbold Carrie, Eastwood Hayden, O'Leary Stephen J. Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2017. DOI: 10.1097/mao.0000000000001589.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.