Comparison of an AI-based mobile pupillometry system and NPi-200 for pupillary light reflex and correlation with glaucoma-related markers.

Jaworski Damian, Suwała Karolina, Kaluzny Bartlomiej J, Kaluzny Jakub J

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Glaucoma is a leading cause of blindness, often progressing asymptomatically until significant vision loss occurs. Early detection is crucial for preventing irreversible damage. The pupillary light reflex (PLR) has proven useful in glaucoma diagnosis, and mobile technologies like the AI-based smartphone pupillometer (AI Pupillometer) offer a promising solution for accessible screening. This study assesses the reliability of the AI Pupillometer in detecting glaucoma.

METHODS

In Experiment 1, 20 healthy participants were assessed using both the AI Pupillometer and the NPi-200 device to evaluate equivalence in measuring PLR. Each eye underwent three trials. Experiment 2 included 46 participants, 24 with primary open-angle glaucoma (POAG) and 22 healthy controls. PLR measurements from the AI Pupillometer were correlated with structural and functional ocular parameters. An additional study expanded the sample to 387 participants (103 glaucoma patients, 284 controls), focusing on differential pupillometry parameters to minimize ambient light interference.

RESULTS

In Experiment 1, the AI Pupillometer demonstrated strong correlations with the NPi-200 in key parameters like initial pupil size (r = 0.700), constricted pupil size (r = 0.755), and constriction velocity (r = 0.541), confirming its reliability. In Experiment 2, although no statistically significant differences in light-corrected PLR parameters were found between groups, glaucoma patients had a marginally higher constricted pupil size (p = 0.1632). Significant correlations were observed between pupillometry and advanced ocular imaging results, notably between constriction amplitude and visual field loss. The additional study revealed significant differences in constriction amplitude (p = 0.014) and relative pupil size change (p = 0.0072) between glaucoma patients and controls, reinforcing the AI Pupillometer's diagnostic potential.

CONCLUSION

This study confirms the AI Pupillometer as a reliable, accessible tool for glaucoma screening. Mobile diagnostics could enhance early detection, improving outcomes for glaucoma patients.

Для цитирования: Jaworski Damian, Suwała Karolina, Kaluzny Bartlomiej J, Kaluzny Jakub J. Comparison of an AI-based mobile pupillometry system and NPi-200 for pupillary light reflex and correlation with glaucoma-related markers.. Frontiers in neurology. 2025. DOI: 10.3389/fneur.2024.1426205.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.