Chauhan Indrajeet, Swami Himanshu, Natraj Rashmi
Аннотация
BACKGROUND: Our study was a prospective, non-randomised, single tertiary care centre study involving 50 children below 10 years, all implanted with Nucleus 24™ cochlear implants to find the relationship between intra-operative NRT (Neural Response Telemetry) values and post-operatively behaviourally obtained Threshold (T) and Comfort (C) levels.
METHODS: NRT threshold values were obtained intra-operatively by using Custom Sound EP Software V 4.4™ (Cochlear Corporation). At switch-on after three weeks, behavioural T and C levels were measured based on behavioural responses given by patients using Custom Sound Version 4.4 software™ (Cochlear Corporation). NRT values were also measured at switch-on, 2 months, 3 months and 6 months.
RESULTS: Intra-operative NRT levels (187.96 ± 12.48) were higher than both T (160.63 ± 22.69) and C (181.21 ± 22.41) levels obtained after 3 weeks. In addition, there was a weak correlation of NRT values with T (P value: 0.05, r = 0.391) and C (P value: 0.05, r = 0.390) levels.
CONCLUSION: NRT is a quick and non-invasive tool to confirm cochlear implant integrity. There is a weak correlation of NRT values with both T and C levels and therefore, intra-operative NRT is a weak predictor for setting both T and C levels at switch-on.
METHODS: NRT threshold values were obtained intra-operatively by using Custom Sound EP Software V 4.4™ (Cochlear Corporation). At switch-on after three weeks, behavioural T and C levels were measured based on behavioural responses given by patients using Custom Sound Version 4.4 software™ (Cochlear Corporation). NRT values were also measured at switch-on, 2 months, 3 months and 6 months.
RESULTS: Intra-operative NRT levels (187.96 ± 12.48) were higher than both T (160.63 ± 22.69) and C (181.21 ± 22.41) levels obtained after 3 weeks. In addition, there was a weak correlation of NRT values with T (P value: 0.05, r = 0.391) and C (P value: 0.05, r = 0.390) levels.
CONCLUSION: NRT is a quick and non-invasive tool to confirm cochlear implant integrity. There is a weak correlation of NRT values with both T and C levels and therefore, intra-operative NRT is a weak predictor for setting both T and C levels at switch-on.