MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.mjafi.2020.10.011

Медицинская библиотека · Medical journal, Armed Forces India · 2020

A study on relationship between Neural Response Telemetry and behavioural Threshold/Comfort levels in children with cochlear implant.

Chauhan Indrajeet, Swami Himanshu, Natraj Rashmi

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Our study was a prospective, non-randomised, single tertiary care centre study involving 50 children below 10 years, all implanted with Nucleus 24™ cochlear implants to find the relationship between intra-operative NRT (Neural Response Telemetry) values and post-operatively behaviourally obtained Threshold (T) and Comfort (C) levels.

METHODS

NRT threshold values were obtained intra-operatively by using Custom Sound EP Software V 4.4™ (Cochlear Corporation). At switch-on after three weeks, behavioural T and C levels were measured based on behavioural responses given by patients using Custom Sound Version 4.4 software™ (Cochlear Corporation). NRT values were also measured at switch-on, 2 months, 3 months and 6 months.

RESULTS

Intra-operative NRT levels (187.96 ± 12.48) were higher than both T (160.63 ± 22.69) and C (181.21 ± 22.41) levels obtained after 3 weeks. In addition, there was a weak correlation of NRT values with T (P value: 0.05, r = 0.391) and C (P value: 0.05, r = 0.390) levels.

CONCLUSION

NRT is a quick and non-invasive tool to confirm cochlear implant integrity. There is a weak correlation of NRT values with both T and C levels and therefore, intra-operative NRT is a weak predictor for setting both T and C levels at switch-on.

Для цитирования: Chauhan Indrajeet, Swami Himanshu, Natraj Rashmi. A study on relationship between Neural Response Telemetry and behavioural Threshold/Comfort levels in children with cochlear implant.. Medical journal, Armed Forces India. 2020. DOI: 10.1016/j.mjafi.2020.10.011.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.