Bicuspid Aortic Valve Stenosis: Favorable Early Outcomes With a Next-Generation Transcatheter Heart Valve in a Multicenter Study.

Perlman Gidon Y, Blanke Philipp, Dvir Danny, Pache Gregor, Modine Thomas, Barbanti Marco, Holy Erik W, Treede Hendrik, Ruile Philipp, Neumann Franz-Josef, Gandolfo Caterina, Saia Francesco, Tamburino Corrado, Mak George, Thompson Christopher, Wood David, Leipsic Jonathon, Webb John G

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study evaluated the results of transcatheter aortic valve replacement (TAVR) in bicuspid aortic stenosis (AS) using a new-generation TAVR device.

BACKGROUND

A bicuspid AS is often considered a relative contraindication to TAVR. Although initial reports have demonstrated feasibility using early-generation devices, outcomes have not matched those seen with tricuspid AS. Paravalvular aortic regurgitation (AR) has been particularly problematic.

METHODS

We collected baseline characteristics, procedural data, and 30-day clinical follow-up findings from 8 centers in Europe and Canada that had performed TAVR in bicuspid AS using the SAPIEN 3 valve.

RESULTS

51 patients underwent TAVR using the SAPIEN 3 valve. Patient mean age was 76.2 ± 9.3 years and the Society of Thoracic Surgeons predicted risk of mortality scores were 5.2 ± 3.7%. Bicuspid valve types were: type 0, 11.8%; type 1, 82.3%; and type 2, 1.9%. There were no cases of valve embolization or need for a second valve. Post-dilation was performed in 7.8%. The mean aortic gradient decreased from 49.4 ± 16.0 mm Hg to 11.2 ± 4.7 mm Hg. Post-implantation AR was none/trivial in 63% and mild in 37%. There were no cases of moderate or severe AR. At 30-day follow-up, there were 2 deaths (3.9%), 2 major vascular complications, and 12 patients (23.5%) required pacemaker implantation.

CONCLUSIONS

TAVR in bicuspid AS using a new-generation device was feasible and effective with favorable valve performance and no cases of moderate or severe AR.

Для цитирования: Perlman Gidon Y, Blanke Philipp, Dvir Danny, Pache Gregor, Modine Thomas, Barbanti Marco, Holy Erik W, Treede Hendrik, Ruile Philipp, Neumann Franz-Josef, Gandolfo Caterina, Saia Francesco, Tamburino Corrado, Mak George, Thompson Christopher, Wood David, Leipsic Jonathon, Webb John G. Bicuspid Aortic Valve Stenosis: Favorable Early Outcomes With a Next-Generation Transcatheter Heart Valve in a Multicenter Study.. JACC. Cardiovascular interventions. 2016. DOI: 10.1016/j.jcin.2016.01.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.