DOI 10.3390/diagnostics16040526

Медицинская библиотека · Diagnostics (Basel, Switzerland) · 2026

Clinical Comparative Evaluation of Two New-Generation Optical Biometers: Intrasession Repeatability and Agreement of Biometric Parameters.

Bedrán Farid J, Piñero David P

СтатьяПеревод пока не подготовлен

Аннотация

Objectives

This study aims to evaluate the intrasession repeatability of the HBM-1 (Huvitz) optical biometer and to assess its agreement with the IOLMaster 700 (Zeiss) for the main biometric parameters used in intraocular lens (IOL) power calculation and in myopia management.

Methods

A cross-sectional observational study was conducted in 82 eyes of 82 patients with the following age distribution: pediatric 26.8%, young adults 35.4%, and older adults 37.8% (total range 6-79 years). Optical biometry was performed three consecutive times with both biometers in the same session by a single trained examiner. HBM-1 repeatability was assessed using the within-subject standard deviation (Sw), the repeatability coefficient, and the intraclass correlation coefficient (ICC). Agreement between biometers was analyzed using Bland-Altman plots (limits of agreement, LoA).

Results

The HBM-1 showed excellent intrasession repeatability, with very low Sw values-on the order of hundredths of a millimeter for axial length (AL), anterior chamber depth (ACD), and lens thickness (LT), and hundredths of a diopter for keratometry-with ICC ≥ 0.97 for most parameters. The mean bias (HBM-1 vs. IOLMaster 700) was small: AL 0.012 ± 0.052 mm (p = 0.045; LoA: -0.09 to 0.11 mm), ACD 0.059 ± 0.068 mm (p < 0.001; -0.07 to 0.19 mm), LT 0.052 ± 0.090 mm (p < 0.001; -0.12 to 0.23 mm), and central corneal thickness 0.82 ± 7.12 μm (p = 0.301; -13.1 to 14.8 μm). For corneal diameter and corneal curvature, mean differences were small (≤0.07 D) and not statistically significant in most cases. Age was not associated with discrepancies in AL but showed weak correlations with some anterior segment differences, without clear clinical relevance.

Conclusions

The HBM-1 demonstrated excellent intrasession repeatability and a good level of clinical agreement with the IOLMaster 700 in a broad population that included children, young adults, and older adults.

Для цитирования: Bedrán Farid J, Piñero David P. Clinical Comparative Evaluation of Two New-Generation Optical Biometers: Intrasession Repeatability and Agreement of Biometric Parameters.. Diagnostics (Basel, Switzerland). 2026. DOI: 10.3390/diagnostics16040526.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.