Cardiac function evaluation in healthy volunteers and patients with implantable cardioverter-defibrillators using high-bandwidth spoiled gradient-echo cine.

Sheagren Calder D, Shadafny Naseem, Escartin Terenz, Terricabras Maria, Cheung Christopher C, Roifman Idan, Wright Graham A

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Implantable cardioverter-defibrillators (ICDs) cause banding artifacts around areas of B0 inhomogeneity in conventional steady-state free precession (SSFP) cine sequences. Alternatively, high-bandwidth gradient-recalled echo (GRE) cine sequences can be used to minimize artifacts in the myocardium. In this study, we assessed the bias and interobserver variability in cardiac volumes and ejection fractions between GRE cines in acquired in the presence of ICDS and ground-truth SSFP cines (without ICDs present) in a population of healthy volunteers. Further, a small cohort of ICD patients was recruited and scanned to demonstrate clinical feasibility.

METHODS

High-bandwidth GRE cine was performed in 11 healthy volunteers with taped ICDs mimicking clinical implants. After the ICD was removed, ground-truth SSFP cine was performed. Two observers separately assessed image quality metrics and contoured the cine images to return cardiac volumes and ejection fractions. Nine patients with an ICD were also scanned with the GRE cine protocol before contrast administration; data were contoured by two observers and analyzed for interobserver agreement.

RESULTS

In the healthy volunteer dataset, no statistically significant differences were found when comparing volumes or ejection fractions between sequences (p > 0.05). Statistically significant differences were found when comparing right ventricular ejection fraction (RVEF) (p = 0.009) and right ventricular end-systolic volume (p = 0.029) between observers, with no other significant interobserver differences. The interobserver variability of patient left ventricular ejection fraction and RVEF data was 3-4%, with lower image quality metrics for patient scans than volunteer scans.

CONCLUSION

GRE cine imaging in healthy volunteers with taped ICDs demonstrated good agreement with SSFP cine, but increased interobserver variability. In patients, reducing the breath-hold duration caused a decrease in image quality, with GRE cine imaging in patients with ICDs demonstrating poorer image quality and greater interobserver variability than in healthy volunteer studies. Future work is needed to improve GRE cine image quality in patients with ICDs to reduce interobserver variability and improve clinical confidence.

Для цитирования: Sheagren Calder D, Shadafny Naseem, Escartin Terenz, Terricabras Maria, Cheung Christopher C, Roifman Idan, Wright Graham A. Cardiac function evaluation in healthy volunteers and patients with implantable cardioverter-defibrillators using high-bandwidth spoiled gradient-echo cine.. Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance. 2025. DOI: 10.1016/j.jocmr.2025.101893.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.