Biometry combining Scheimpflug imaging and partial coherence interferometry versus standard partial coherence interferometry biometry.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
van Renterghem V E W C M, Crahay F-X
Аннотация
PURPOSE: To compare measurements of mean keratometry (K), axial length (AL), Jackson first and second astigmatism vector (J0 and J45), astigmatism magnitude, white-to-white (WTW) diameter and anterior chamber depth (ACD) values between two biometers, the Pentacam® AXL and the IOL Master 500.
SETTING: Public hospital in Belgium.
DESIGN: Retrospective instrument reliability analysis.
METHODS: Eyes of cataract or refractive lens exchange patients were examined preoperatively on a device combining Scheimpflug imaging and partial coherence interferometry (Pentacam® AXL) and a reference optical biometer (IOL Master 500). K, AL, J0, J45, astigmatism, WTW and ACD values were compared.
RESULTS: Three hundred and twenty-eight eyes of 187 cataract or refractive lens exchange patients (mean age 68years) were examined preoperatively. Measurements of mean keratometry, axial length, ACD, white-to-white and astigmatism were significantly different (P<0.05) between the two devices. Measurements of J0 and J45 did not significantly differ. Keratometry measurements with the Pentacam® AXL were significantly flatter than those measured with the IOL Master 500. Axial length measured with Pentacam® AXL was significantly greater when measured with the IOL Master 500, but this difference was not clinically relevant (0.015mm).
CONCLUSIONS: The Pentacam® AXL and the IOL Master 500 cannot be used interchangeably.
SETTING: Public hospital in Belgium.
DESIGN: Retrospective instrument reliability analysis.
METHODS: Eyes of cataract or refractive lens exchange patients were examined preoperatively on a device combining Scheimpflug imaging and partial coherence interferometry (Pentacam® AXL) and a reference optical biometer (IOL Master 500). K, AL, J0, J45, astigmatism, WTW and ACD values were compared.
RESULTS: Three hundred and twenty-eight eyes of 187 cataract or refractive lens exchange patients (mean age 68years) were examined preoperatively. Measurements of mean keratometry, axial length, ACD, white-to-white and astigmatism were significantly different (P<0.05) between the two devices. Measurements of J0 and J45 did not significantly differ. Keratometry measurements with the Pentacam® AXL were significantly flatter than those measured with the IOL Master 500. Axial length measured with Pentacam® AXL was significantly greater when measured with the IOL Master 500, but this difference was not clinically relevant (0.015mm).
CONCLUSIONS: The Pentacam® AXL and the IOL Master 500 cannot be used interchangeably.