BMC ophthalmology · 2024

Agreement between a new fully automatic ocular biometer based on optical low-coherence reflectometry and an optical biometer based on Scheimpflug imaging combined with partial coherence interferometry.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ning Rui, Xu Huilin, Li Zheng, Yu Jinjin, Xu Shuoyu, Lei Chak Seng, Wang Yiran, Savini Giacomo, Schiano-Lomoriello Domenico, Zhou Xingtao, Huang Jinhai

Аннотация

BACKGROUND: To analyze the difference and agreement between measurements obtained by a new fully automatic optical biometer, the SW-9000 μm Plus, based on optical low-coherence reflectometry (OLCR) and a commonly used optical biometer (Pentacam AXL) based on Scheimpflug imaging with partial coherence interferometry (PCI).

METHODS: The central corneal thickness (CCT), anterior chamber depth (ACD, from epithelium to anterior lens surface), lens thickness (LT), mean keratometry (Km), corneal astigmatism, corneal diameter (CD), pupil diameter (PD), and axial length (AL) of 74 eyes (from 74 healthy subjects) were measured using the SW-9000 μm Plus and the Pentacam AXL to determine the agreement. Double angle plots were used for astigmatism vector analysis. Bland-Altman and 95% limits of agreement (LoA) were calculated.

RESULTS: Statistically significant differences were detected for all parameters but J0 vector. The Bland-Altman analysis of AL, CCT, ACD, Km, CD, J0 and J45 indicated a high level of agreement between the two devices. Among AL, CCT, ACD, Km, J0, J45, CD, and PD, the 95% LoA ranged from  -0.07 to 0.05 mm, -9.67 to 7.34 mm, -0.11 to 0.04 mm, -0.25 to 0.50 D, -0.22 to 0.20 D, -0.15 to 0.20 D, -0.23 to 0.35 mm and 1.55 to 3.77 mm, respectively.

CONCLUSIONS: The measurements of AL, CCT, ACD, Km, corneal astigmatism, and CD showed a narrow LoA and may be used interchangeably in healthy subjects between the new OLCR optical biometer and the Scheimpflug/PCI biometer; however, a poor agreement was noted for PD values.