MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41598-024-76031-7

Медицинская библиотека · Scientific reports · 2024

Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes.

Lee Yunjin, Oh Joo Youn, Choi Hyuk Jin, Kim Mee Kum, Yoon Chang Ho

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the agreement of ocular biometry measured using a swept-source optical coherence (Casia 2) and a dual Scheimpflug (Galilei G6) tomography in keratoconus.

METHODS

This retrospective study included 102 eyes from 102 keratoconus patient examined using both devices. Parameters compared included flat (Kf) and steep (Ks) keratometry, astigmatism of anterior, posterior, and total keratometry, central (CCT) and thinnest (TCT) corneal thickness. To assess the agreement, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used.

RESULTS

Anterior and total Ks and Kf showed moderate or good agreement with 95% limits of agreement (LoA) range over 9.75 D. Posterior Ks and Kf were lower in the Galilei G6 (all p < 0.001). Astigmatism showed moderate, poor, and moderate agreement for anterior, posterior, and total keratometry, respectively. CCT and TCT showed excellent agreement; however, the 95% LoA range was over 60 μm.

CONCLUSION

The agreement between the two devices was not excellent for most parameters used to diagnose keratoconus and assess disease progression, and the differences were clinically significant. Therefore, the measurements from these two devices are not interchangeable for patients with keratoconus.

Для цитирования: Lee Yunjin, Oh Joo Youn, Choi Hyuk Jin, Kim Mee Kum, Yoon Chang Ho. Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes.. Scientific reports. 2024. DOI: 10.1038/s41598-024-76031-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.