DOI 10.1136/bmjophth-2024-001875

Медицинская библиотека · BMJ open ophthalmology · 2024

Comparison of two digital alignment systems for toric intraocular lens implantation.

Zeilinger Johannes, Schlatter Andreas, Ruiss Manuel, Bayer Natascha, Kronschläger Martin, Findl Oliver

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the two most used digital alignment systems regarding precision, repeatability and loss of track.

METHODS

15 eyes of 15 patients older than 21 years with cataracts were included in this prospective study. The two systems were intraoperatively superimposed and recorded, and the alignment of the two displayed alignment axes was analysed regarding precision, repeatability and loss of track.

RESULTS

There was a significant difference in precision and repeatability between the two digital alignment systems regarding the projected alignment axis. The deviation from the actual target axis was significantly different, with a mean of 0.34°±0.75° for the Zeiss system and 1.60°±1.08° for the Alcon system (p=0.03, n=14). The within-subject SD was significantly lower with 0.21° for the Zeiss system and 0.34° for the Alcon system (p=0.03, n=14).

CONCLUSIONS

The Zeiss Callisto system showed a significantly lower deviation from the target axis, higher stability with eye movements and less need for microscope illumination than the Alcon system. Both systems showed high precision when compared with manual marking methods.

TRIAL REGISTRATION NUMBER: NCT05220683.

Для цитирования: Zeilinger Johannes, Schlatter Andreas, Ruiss Manuel, Bayer Natascha, Kronschläger Martin, Findl Oliver. Comparison of two digital alignment systems for toric intraocular lens implantation.. BMJ open ophthalmology. 2024. DOI: 10.1136/bmjophth-2024-001875.