Obaid Daniel R, Smith David, Gilbert Michael, Ashraf Saeed, Chase Alexander
Аннотация
BACKGROUND: We evaluated the utility of the HeartNavigator III software (Philips Healthcare, Netherlands) to create a patient specific three-dimensional model using ECG-gated CT images to plan Transcatheter Aortic Valve Replacement (TAVR) in a patient with previous Starr-Edwards mitral prosthesis.
METHODS: A patient with a previous Starr-Edwards mitral prosthesis considered too high risk for conventional surgery required TAVR. It was uncertain whether this would be possible whilst avoiding the complication of the aortic prosthesis interacting with the high-profile Starr-Edwards cage and ball valve mechanism. To ensure it would be feasible and aid in the planning of the procedure a patient specific three-dimensional model was created from ECG-gated CT images using HeartNavigator III software (Philips Healthcare, Netherlands).
RESULTS: The patient specific model allowed simulated "virtual" TAVR implantations to be performed with different models and sizes of aortic prosthesis. These pre-implant simulations allowed a safe and feasible implant strategy to be chosen. The images were also co-registered with fluoroscopy to guide deployment.
CONCLUSION: Using a patient-specific CT simulation technique we performed TAVR with a high level of precision, achieving a clear margin between a Portico (Abbot Vascular, US) TAVR and the Starr-Edwards cage.
METHODS: A patient with a previous Starr-Edwards mitral prosthesis considered too high risk for conventional surgery required TAVR. It was uncertain whether this would be possible whilst avoiding the complication of the aortic prosthesis interacting with the high-profile Starr-Edwards cage and ball valve mechanism. To ensure it would be feasible and aid in the planning of the procedure a patient specific three-dimensional model was created from ECG-gated CT images using HeartNavigator III software (Philips Healthcare, Netherlands).
RESULTS: The patient specific model allowed simulated "virtual" TAVR implantations to be performed with different models and sizes of aortic prosthesis. These pre-implant simulations allowed a safe and feasible implant strategy to be chosen. The images were also co-registered with fluoroscopy to guide deployment.
CONCLUSION: Using a patient-specific CT simulation technique we performed TAVR with a high level of precision, achieving a clear margin between a Portico (Abbot Vascular, US) TAVR and the Starr-Edwards cage.
Предметные рубрики MeSH
Aged, 80 and overAortic ValveAortic Valve StenosisCardiac-Gated Imaging TechniquesEchocardiography, TransesophagealElectrocardiographyFemaleHeart Valve ProsthesisHeart Valve Prosthesis ImplantationHumansImaging, Three-DimensionalMitral ValveMultidetector Computed TomographyPatient-Specific ModelingPredictive Value of TestsProsthesis DesignRadiographic Image Interpretation, Computer-AssistedSeverity of Illness IndexSurgery, Computer-AssistedTranscatheter Aortic Valve ReplacementTreatment Outcome