DOI 10.1109/tbme.2026.3668134

Медицинская библиотека · IEEE transactions on bio-medical engineering · 2026

Tunable Magnetically Actuated Retraction Device for Improved Control During Endoscopic Tissue Manipulation.

Mayer Haley, Shlomovitz Eran, Drake James, Looi Thomas, Diller Eric

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Endoscopic submucosal dissection (ESD) is a minimally-invasive method for removing gastrointestinal lesions. Adoption has been slow due to the procedure's technical difficulty, length of surgery, and lack of appropriate tools. Tissue traction is required to reduce the technical difficulty of ESD and ensure safe lesion removal. This paper presents the first ESD retraction device that is wireless, endoscope-independent, and provides adjustable traction force control.

METHODS

The magnetic retraction device is 3.25 mm wide and 45 mm long. It is actuated wirelessly by a rotating permanent magnet held outside the abdomen. The device design is informed by a presented model of actuation, and we show characterization of the device actuation, validation of mechanical retraction, and clinical feasibility through an ex-vivo experiment.

RESULTS

Experiments show that the device retracts 20 mm at 0.2 millimeters per second and generates a peak retraction force of 7.98 N and a four-peak average of 5.62 N at an external magnet-to-device distance of 55 mm. Ex-vivo testing with a gastric porcine model was performed with a dissection speed of 11.3 square centimeters per hour, 1.25 times faster than the reported international benchmark.

CONCLUSION

This enhanced tissue control would greatly improve patient clinical outcomes by expanding the use of ESD, shortening procedure times, minimizing complications, and potentially further expand its future medical applications.

SIGNIFICANCE: This work grants surgeons improved tissue control throughout an ESD procedure, providing adequate visualization of the dissection plane, control independent of the flexible endoscope, and deployment anywhere along the gastrointestinal tract.

Для цитирования: Mayer Haley, Shlomovitz Eran, Drake James, Looi Thomas, Diller Eric. Tunable Magnetically Actuated Retraction Device for Improved Control During Endoscopic Tissue Manipulation.. IEEE transactions on bio-medical engineering. 2026. DOI: 10.1109/tbme.2026.3668134.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.