m+Medelevator.RU
← Поиск публикаций
Magnetic resonance in medicine · 2026

Toward Optimized Intravoxel Incoherent Motion (IVIM) and Compartmental T2 Mapping in Abdominal Organs.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Stabinska Julia, Thiel Thomas A, Wittsack Hans-Jörg, Ljimani Alexandra, Zöllner Helge J

Аннотация

PURPOSE: To quantitatively assess the bias in the intravoxel incoherent motions (IVIM)-derived pseudo-diffusion volume fraction (f) caused by the differences in relaxation times between the tissue and fluid compartments, and to develop a two-dimensional (b-value-TE) fitting approach for simultaneous T2 and IVIM parameter estimation along with an optimal acquisition protocol for the relaxation-compensated T2-IVIM imaging in the liver.

METHODS: Simulations were conducted to investigate the TR- and TE-dependent bias in f when using the IVIM model, and to evaluate the applicability of the 2D T2-IVIM model for reducing this bias. The numerical findings were then validated using the in vivo IVIM data from four healthy volunteers on a 3-Tesla MRI scanner. Finally, a numerical framework for optimizing the T2-IVIM protocol for relaxation-compensated f parameter estimation was proposed and tested using the in vivo data.

RESULTS: In vivo, the traditional IVIM model showed a trend toward higher f with increasing TE in the liver (R = 0.46, p = 0.023), but not in the kidney cortex (R = -0.067, p = 0.76) or medulla (R = 0.039, p = 0.86). In both simulations and in vivo, 2D T2-IVIM modeling yielded lower f values and reduced variability in the liver. Our results further suggest that a b-TE protocol with six b-values and three TEs (50, 60, and 100 ms) may be optimal for liver T2-IVIM.

CONCLUSION: The extended 2D T2-IVIM model effectively minimizes the TE-dependent bias in f and allows simultaneous estimation of the IVIM parameter and compartmental T2 values in abdominal organs.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты