In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs.

Camps Vicente J, Tolosa Angel, Piñero David P, de Fez Dolores, Caballero María T, Miret Juan J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To analyze the "in vitro" aberrometric pattern of a refractive IOL and two extended depth of focus IOLs.

METHODS

A special optical bench with a Shack-Hartmann wavefront sensor (SH) was designed for the measurement. Three presbyopia correction IOLs were analyzed: Mini WELL (MW), TECNIS Symfony ZXR00 (SYM), and Lentis Mplus X LS-313 MF30 (MP). Three different pupil sizes were used for the comparison: 3, 4, and 4.7 mm.

RESULTS

MW generated negative primary and positive secondary spherical aberrations (SA) for the apertures of 3 mm (-0.13 and +0.12 μm), 4 mm (-0.12 and +0.08 μm), and 4.7 mm (-0.11 and +0.08 μm), while the SYM only generated negative primary SA for 4 and 4.7 mm apertures (-0.12 μm and -0.20 μm, resp.). The MP induced coma and trefoil for all pupils and showed significant HOAs for apertures of 4 and 4.7 mm.

CONCLUSIONS

In an optical bench, the MW induces negative primary and positive secondary SA for all pupils. The SYM aberrations seem to be pupil dependent; it does not produce negative primary SA for 3 mm but increases for higher pupils. Meanwhile, the HOAs for the MW and SYM were not significant. The MP showed in all cases the highest HOAs.

Для цитирования: Camps Vicente J, Tolosa Angel, Piñero David P, de Fez Dolores, Caballero María T, Miret Juan J. In Vitro Aberrometric Assessment of a Multifocal Intraocular Lens and Two Extended Depth of Focus IOLs.. Journal of ophthalmology. 2017. DOI: 10.1155/2017/7095734.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.