Changes in Keratometric Values and Corneal High Order Aberrations After Hydrogel Inlay Implantation.

Whang Woong-Joo, Yoo Young-Sik, Joo Choun-Ki, Yoon Geunyoung

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: We sought to analyze surgically induced refractive change (SIRC) and change in high-order aberration after Raindrop corneal inlay insertion (ReVision Optics, Lake Forest, CA), and assess the extent to which Raindrop corneal inlay insertion could correct presbyopia.

DESIGN

Interventional case series.

METHODS

Seventeen patients were included if they had a corneal thickness ≥500 μm and a stable manifest spherical equivalent refraction between 0.50 and +1.00 diopters (D). The Raindrop corneal inlay was implanted on the stromal bed of a femtosecond laser-assisted generated flap of nondominant eyes. Manifest refraction, corneal powers, and corneal high-order aberrations were measured preoperatively and at 3 and 12 months postoperatively.

RESULTS

The SIRC by manifest refraction was 0.99 ± 0.26 D. The changes derived from simulated keratometry (K), true net power, and equivalent K reading (EKR) at 1.0-4.0 mm were greater than the SIRC (all P < .01) while the change in EKR at 6.0 mm was less than the SIRC (P < .01). The changes in EKR 5.0 mm, automated K, and EKR 4.5 mm did not differ significantly from the SIRC (P = .81, .29, and .09, respectively), and the difference was the least for EKR 5.0 mm. In analysis of high-order aberrations, only spherical aberration showed statistically significant difference between preoperative and postoperative on both anterior cornea and total cornea (all P < .01).

CONCLUSION

Raindrop corneal inlay corrects presbyopia via increasing negative spherical aberration. The equivalent K reading at 5.0 mm accurately reflected the SIRC, and would be applicable for intraocular power prediction before cataract surgery.

Для цитирования: Whang Woong-Joo, Yoo Young-Sik, Joo Choun-Ki, Yoon Geunyoung. Changes in Keratometric Values and Corneal High Order Aberrations After Hydrogel Inlay Implantation.. American journal of ophthalmology. 2016. DOI: 10.1016/j.ajo.2016.09.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.