Implementation of an Intraoperative Augmented Reality Environment for Custom-Made Partial Pelvis Replacements-A Proof of Concept and Initial Results.
Аннотация
Background
The use of augmented reality (AR) in orthopaedics is growing rapidly but is mainly limited to pre-operative planning and teaching. This study is one of the first to describe the intraoperative application within revision arthroplasty for the positioning of customised partial pelvic replacements.
Methods
In a proof-of-concept study an AR environment was used during surgery in 11 cases to enhance implant positioning. Postoperatively, a voxel-based CT deviation analysis was carried out to determine the COR deviation and the cup plane deviation angle. Additionally, digital implant superimposition was conducted.
Results
Implantation was possible in all cases with a mean COR deviation vector of 4.2 (SD 2.5; 1.2-9.3) mm and a cup plane deviation angle of 4.4 (SD 2.5; 0.7-8.1)°. The implant analysis showed a superimposition of 0.69 (SD 0.15; 0.38-0.88) (Dice-Score calculation).
Conclusions
This study is able to report promising results for AR in orthopaedic surgery, showing improved intraoperative feedback in complex operations, resulting in increased accuracy. However, the integration of AR poses a new challenge to the surgical team, especially because the AR users are facing a significantly increased level of intraoperative stress. Further development of this auspicious tool, as well as a conceivable combination with navigation, is necessary to facilitate broader usage.