MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.prro.2014.03.010

Медицинская библиотека · Practical radiation oncology · 2014

Effect of metal hip prosthesis on the accuracy of electromagnetic localization tracking.

Bittner Nathan, Butler Wayne M, Kurko Brian S, Merrick Gregory S

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To quantify the effect of metal hip prosthesis on the ability to track and localize electromagnetic transponders.

METHODS AND MATERIALS: Three Calypso Beacon (Varian Medical Systems, Palo Alto, CA) transponders were implanted into 2 prostate phantoms. The geometric center of the transponders were identified on computed tomography and set as the isocenter. With the phantom stationary on the treatment table and the tracking array 14-cm above the isocenter, data were acquired by the Calypso system at 10 Hz to establish the uncertainty in measurements. Transponder positional data were acquired with unilateral hip prostheses of different metallic compositions and then with bilateral hips placed at variable separation from the phantom.

RESULTS

Regardless of hip prosthesis composition, the average vector displacement in the presence of a unilateral prosthesis was <0.5 mm. The greatest contribution to overall vector displacement occurred in the lateral dimension. With bilateral hip prosthesis, the average vector displacement was 0.3 mm. The displacement in the lateral dimension was markedly reduced compared with a unilateral hip, suggesting that there was a countervailing effect with bilateral hip prosthesis. The greatest average vector displacement was 0.6 mm and occurred when bilateral hip prostheses were placed within 4 cm of the detector array.

CONCLUSIONS

Unilateral and bilateral hip prostheses did not have any meaningful effect on the ability to accurately track electromagnetic transponders implanted in a prostate phantom. At clinically realistic distances between the hip and detection array, the average tracking error is negligible.

Для цитирования: Bittner Nathan, Butler Wayne M, Kurko Brian S, Merrick Gregory S. Effect of metal hip prosthesis on the accuracy of electromagnetic localization tracking.. Practical radiation oncology. 2014. DOI: 10.1016/j.prro.2014.03.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.