Diagnosis of feline filariasis assisted by a novel semi-automated microfluidic device in combination with high resolution melting real-time PCR.

Phuakrod Achinya, Sripumkhai Witsaroot, Jeamsaksiri Wutthinan, Pattamang Pattaraluck, Juntasaro Ekachai, Thienthong Therdthai, Foongladda Suporn, Brindley Paul J, Wongkamchai Sirichit

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The diagnosis of filariasis traditionally relies on the detection of circulating microfilariae (mf) using Giemsa-stained thick blood smears. This approach has several limitations. We developed a semi-automated microfluidic device to improve and simplify the detection of filarial nematodes.

METHODS

The efficiency and repeatability of the microfluidic device was evaluated. Human EDTA blood samples were 'spiked' with B. malayi mf at high, moderate, and low levels, and subsequently tested 10 times. The device was also used for a field survey of feline filariasis in 383 domesticated cats in an area of Narathiwat Province, Thailand, the endemic area of Brugia malayi infection.

RESULTS

In the control blood arbitrarily spiked with mf, the high level, moderate level and low level mf-positive controls yielded coefficient variation (CV) values of 4.44, 4.16 and 4.66%, respectively, at the optimized flow rate of 6 µl/min. During the field survey of feline filariasis in Narathiwat Province, the device detected mf in the blood of 34 of 383 cats (8.9%) whereas mf were detected in 28 (7.3%) cats using the blood smear test. Genomic DNA was extracted from mf trapped in the device after which high-resolution melting (HRM) real-time PCR assay was carried out, which enabled the simultaneous diagnosis of filarial species. Among the 34 mf-positive samples, 12 were identified as B. malayi, 15 as Dirofilaria immitis and 7 as| D. repens.

CONCLUSIONS

We developed a semi-automated microfluidic device to detect mf of filarial parasites that could be used to diagnose lymphatic filariasis in human populations. This novel device facilitates rapid, higher-throughput detection and identification of infection with filariae in blood samples.

Для цитирования: Phuakrod Achinya, Sripumkhai Witsaroot, Jeamsaksiri Wutthinan, Pattamang Pattaraluck, Juntasaro Ekachai, Thienthong Therdthai, Foongladda Suporn, Brindley Paul J, Wongkamchai Sirichit. Diagnosis of feline filariasis assisted by a novel semi-automated microfluidic device in combination with high resolution melting real-time PCR.. Parasites & vectors. 2019. DOI: 10.1186/s13071-019-3421-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.