BACKGROUND

Beginning in 2008, metal-on-metal prostheses have been in the spotlight owing to much higher revision rates than expected. Adverse local tissue reactions have been well described in the literature as potential complications.

METHODS

Between 2012 and 2013, 13 patients with metal-on-metal total hip replacements were evaluated clinically and radiologically and with laboratory samples. The same tests were repeated between 2015 and 2016 on eight patients to assess any changes. In the laboratory assessment, we searched for chromium, cobalt, molybdenum, and nickel in blood and urine samples over 24 h.

RESULTS

Clinical assessment has shown good score in all patients except one. On a second examination, between 2015 and 2016, all patients obtained results similar to those obtained in the first assessment, except a patient, who reported a recent fall. In the radiological assessment between 2012 and 2013, results were optimal, apart from a case of aseptic mobilization. The patients reassessed 3 years after the first examination showed radiological results similar to those previously obtained, apart from a patient, who showed signals of mobilization. Metal levels found in their blood decreased in most cases after 3 years. Urine levels of nickel increased in five subjects, and chromium levels increased in four, but levels of cobalt and molybdenum decreased in four patients.

CONCLUSION

It could be hypothesized that the decreasing trend of metal ion levels is associated with a stable wear status. On the contrary, a progressive increase in metal ion levels must be considered as early proof of implant loosening.

Для цитирования: Sessa Giuseppe, Testa Gianluca, Gioitta Iachino Salvatore, Costarella Luciano, Puma Pagliarello Calogero, Ferrante Margherita, Grasso Alfina, Pavone Vito. Potential presence of metals in patients treated with metal-metal coupling prostheses for hip arthroplasty at 7 and 10 years of follow-up.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2018. DOI: 10.1007/s00590-018-2129-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.