Refractive prediction error in cataract surgery using an optical biometer equipped with anterior segment OCT.

Kato Yukihito, Kojima Takashi, Tamaoki Akeno, Ichikawa Kei, Tamura Kensei, Ichikawa Kazuo

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate refractive error after cataract surgery using an optical biometer equipped with anterior segment optical coherence tomography (AS-OCT).

SETTING

Chukyo Eye Clinic, Nagoya, Japan.

DESIGN

Retrospective observational design.

METHODS

In total, 150 patients with cataract (150 eyes, mean age 73.4 ± 8.2 years, men 76, women 74), who underwent measurement of parameters with the AS-OCT scanners ANTERION (AS-OCTB) and IOLMaster 700 (OCTB) before cataract surgery, were enrolled in the study. Refractive prediction error was compared between the 2 devices using the SRK/T, Haigis, and Barrett Universal II (UII) formulas for intraocular lens (IOL) power calculation.

RESULTS

There were significant differences between AS-OCTB and OCTB in axial length, mean corneal refractive power, anterior chamber depth, lens thickness, and corneal diameter (n = 150). In the SRK/T formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.06 ± 0.46 diopters (D) and 0.02 ± 0.42 D, respectively. In the Haigis formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.23 ± 0.40 D and -0.08 ± 0.35 D, respectively. In the Barrett UII formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.02 ± 0.38 D and 0.11 ± 0.36 D, respectively. AS-OCTB showed significantly larger refractive prediction error toward myopia than OCTB in all 3 formulas (P < .0001).

CONCLUSIONS

The refractive prediction error using AS-OCTB showed a small difference from that using OCTB. While clinically comparable, the 2 methods could drive meaningful differences in IOL selection.

Для цитирования: Kato Yukihito, Kojima Takashi, Tamaoki Akeno, Ichikawa Kei, Tamura Kensei, Ichikawa Kazuo. Refractive prediction error in cataract surgery using an optical biometer equipped with anterior segment OCT.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000000781.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.