Twinkling artifact in color Doppler imaging of the orbit.

Ustymowicz Andrze, Krejza Jaroslaw, Mariak Zofia

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To show an artifact related to color Doppler flow imaging of the orbit.

METHODS

Three patients with strongly reflective structures in the orbit were selected from those routinely referred by clinicians for color Doppler ultrasonography of the orbit. Gray scale and color flow images were obtained with a 7.5-MHz linear array probe for a region with strongly reflective structures. A spectral display was acquired to confirm the presence of blood flow.

RESULTS

One patient had a metallic foreign body just behind the bulb; another had calcification within the irregular mass of phthisis bulbi; and the third had hyperechoic drusen in the periphery of the intraocular melanoma. The color mosaic, suggesting the presence of blood flow, was detected beyond all hyperechoic structures. Close vertical bands with no outer wrapping were detected in the spectrum display, obtained by placing the sample volume on the region of color flow. The artificial color flow was recognized as a color Doppler twinkling artifact.

CONCLUSIONS

The color flow beyond the strongly reflecting structures in the orbit might be mistakenly interpreted as real blood flow if an examiner is not familiar with the artifact. It should prompt further imaging with spectral Doppler ultrasonography.

Для цитирования: Ustymowicz Andrze, Krejza Jaroslaw, Mariak Zofia. Twinkling artifact in color Doppler imaging of the orbit.. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2002. DOI: 10.7863/jum.2002.21.5.559.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.