DOI 10.1016/j.hrthm.2025.02.025
Effects of handheld electro-shockers on cardiac implantable electronic devices.
Аннотация
BACKGROUND
Handheld electro-shockers are commonly used for self-defense and possibly interact with cardiac implantable electronic devices (CIEDs).
OBJECTIVE
This study aimed to evaluate the effects of handheld electro-shockers on CIEDs.
METHODS
Six pacemakers (PMs; 1 VVI-PM, 4 DDD-PM, 1 CRT-PM) and 10 implantable cardioverter-defibrillators (ICDs; 5 VVI-ICD, 1 DDD-ICD, 4 CRT-ICD) from different manufacturers were implanted in a subcutaneous and submuscular location in an isolated section of a porcine chest and connected to an interactive heart simulator. Three electro-shockers (PowerMax, 500,000 V; Electric Guard, 250,000 V; Bikenda, <50,000 V) were successively applied to the porcine chest.
RESULTS
No damage to any of the PMs or ICDs was noted during testing. Subcutaneously, the highest rate of interaction was documented with the PowerMax electro-shocker, with 16 of 16 devices (100%) inhibiting stimulation during electro-shock application and 6 of 10 ICDs (60%) storing a ventricular high-rate episode. Application of the Electric Guard electro-shocker resulted in inhibition of stimulation in 10 of 16 devices (63%) and a stored ventricular high-rate episode in 1 of 10 ICDs (10%). The rate of oversensing was lowest with the Bikenda electro-shocker, with inhibition of stimulation in 2 of 16 devices (13%) and storage of a ventricular high-rate episode in none of the 10 ICDs. There was no relevant difference between subcutaneous and submuscular device placement (risk of interaction, 39% vs 36%; P = .88).
CONCLUSION
Handheld electro-shockers pose a substantial acute risk for interaction with CIEDs. The individual risk depends on electro-shocker voltage, manufacturer, and type of cardiac device.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.