Impeller Position in a Magnetically Levitated Rotodynamic Blood Pump and Its Impact on In-Silico Hemocompatibility.

Grujic Marko, Schoefbeck Rosmarie, Thamsen Bente, Aigner Philipp, Röhrich Michael, Jakubek Stefan, Zimpfer Daniel, Granegger Marcus

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

In magnetically levitated rotodynamic blood pumps (RPBs), the impeller position depends on a balance of electromagnetic and fluid dynamic forces. The aim of this study was to describe the impeller position of the HeartMate 3 over a wide range of operating conditions and assess its potential impact on hemocompatibility. Three-dimensional impeller positions were measured using a transparent HeartMate 3 pump casing, laser distance measurements, and a high-speed camera. Accompanying computational fluid dynamic (CFD) hemocompatibility predictions of a displaced and centered impeller at a typical operating point were compared. Impeller positions vary substantially with different operating points with a maximum axial displacement of 223 µm at 7 L/min and 7,000 rpm and a maximum radial displacement of 145 µm at 0 L/min and 7,000 rpm. In CFD, a displaced impeller had only a minor influence on global pump parameters (<2%) at an operating point of 5 L/min and 6,000 rpm. However, deviations in local flow metrics of up to 9% were observed compared with a centered impeller simulation. We here provide the impeller position of the HeartMate 3 over the full operating range (0-9 L/min, 3,000-7,000 rpm) to support further research, including more extensive CFD simulations.

Другие публикации по теме

Heart-Assist DevicesHydrodynamicsComputer SimulationMaterials Testing
Для цитирования: Grujic Marko, Schoefbeck Rosmarie, Thamsen Bente, Aigner Philipp, Röhrich Michael, Jakubek Stefan, Zimpfer Daniel, Granegger Marcus. Impeller Position in a Magnetically Levitated Rotodynamic Blood Pump and Its Impact on In-Silico Hemocompatibility.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2025. DOI: 10.1097/mat.0000000000002434.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.