Durability of output-dependent QRS transition and left bundle branch capture in left bundle branch area pacing.

Shimeno Kenji, Matsumoto Naoki, Tamura Shota, Matsuo Masanori, Hayashi Yusuke, Abe Yukio, Fukuda Daiju

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Although output-dependent QRS transition is a specific indicator that confirms left bundle branch (LBB) capture during left bundle branch area pacing (LBBAP), its durability remains unclear.

OBJECTIVE

The purpose of this study was to evaluate the presence of output-dependent QRS transition and capture thresholds of the LBB and left ventricular septal myocardium immediately and up to 1 year after the LBBAP procedure.

METHODS

We enrolled 129 patients with successful LBBAP who were available for 1-year follow-up postoperatively. Threshold testing was performed immediately after LBBAP on postoperative day 0 (POD-0) and after 3 days (POD-3), 6 months (POD-180), and 1 year (POD-360).

RESULTS

Output-dependent QRS transition persisted in 64 patients (88%) on POD-360, from among the 73 patients with output-dependent QRS transition on POD-0. In contrast, 55 of 56 patients without QRS transition on POD-0 (98%) did not exhibit QRS transition thereafter. LBB thresholds were slightly elevated on POD-360, albeit without statistical significance, compared with those on POD-0 (1.22 ± 1.00 V vs 1.43 ± 1.29 V at 0.4 ms; P = .26). The LBB thresholds increased by ≥1.5 V in 7 patients (11%). However, in 93% of patients with an LBB threshold of ≤2.5 V on POD-0, LBB capture was maintained at 2.5 V on POD-360. Left ventricular septal thresholds were similar on POD-0 and POD-360 (0.81 ± 0.36 V vs 0.83 ± 0.24 V; P > .99) and did not increase by ≥1.5 V in any patient.

CONCLUSION

Output-dependent QRS transitions were highly reproducible after implantation. Furthermore, LBB thresholds remained stable in most cases during the first postoperative year.

Для цитирования: Shimeno Kenji, Matsumoto Naoki, Tamura Shota, Matsuo Masanori, Hayashi Yusuke, Abe Yukio, Fukuda Daiju. Durability of output-dependent QRS transition and left bundle branch capture in left bundle branch area pacing.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.08.039.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.