m+Medelevator.RU
← Поиск публикаций
Pacing and clinical electrophysiology : PACE · 2023

Inappropriate shock incidence in patients with subcutaneous implantable cardioverter-defibrillators with concomitant cardiac implantable electronic devices: A single-center cohort study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Okazaki Makiko, Sahashi Yuki, Nagase Takahiko, Inoue Kanki, Sekiguchi Yukio, Nitta Junichi, Shinoda Satoru, Shimizu Sayuri, Kuroki Makoto, Isobe Mitsuaki, Mihara Takahiro

Аннотация

BACKGROUND: Subcutaneous implantable cardioverter defibrillators (S-ICDs) are occasionally used in combination with other cardiac implantable electronic devices (CIEDs). However, whether the incidence of inappropriate shock increases in patients with S-ICDs and concomitant CIEDs remains unclear. This study aimed to investigate the association between the concomitant use of CIEDs and the incidence of inappropriate shock in patients with current-generation S-ICDs.

METHODS: A total of 127 consecutive patients received an S-ICD. Patients were assigned to two groups depending on concomitant use of CIEDs at the time of S-ICD implantation: patients without other CIEDs (non-combined group, 106 patients) and patients with other CIEDs (combined group, 21 patients). CIEDs included pacemakers, implantable cardioverter defibrillators, cardiac resynchronization therapy pacemakers, and cardiac resynchronization therapy defibrillators. The primary outcome was inappropriate shock, defined as a shock other than ventricular arrhythmia. Hazard ratios and 95% confidence intervals were calculated using a time-varying Cox proportional hazards model which was adjusted for age because age differed between the groups and could be a confounder.

RESULTS: During a median follow-up period of 2.2 years (interquartile range, 1.0-3.4 years), inappropriate shock events occurred in 17 (16%) and five (19%) patients of the non-combined and combined groups, respectively. While the age-adjusted hazard ratio for inappropriate shock was 24% higher in the combined than in the non-combined group (hazard ratio = 1.24, 95% confidence interval, 0.39-3.97), this difference was insignificant (p = .71).

CONCLUSION: The incidence of inappropriate shock did not differ between patients with and without concomitant use of CIEDs, suggesting that S-ICDs could potentially be combined with other CIEDs without increasing the number of inappropriate shocks. Further studies are warranted to confirm the safety and feasibility of concomitant use of S-ICDs and CIEDs.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты