BACKGROUND

In patients with left ventricular assist devices (LVADs), ischemic and hemorrhagic stroke are dreaded complications. Predictive markers for these events are lacking. This study aimed to investigate the prevalence and predictive value of microembolic signals (MES) for stroke, detected by Transcranial Doppler sonography (TCD) in patients with HeartMate 3 (HM 3) or HeartWare (HW).

METHODS

A thirty-minute bilateral TCD monitoring of the middle cerebral artery (MCA) was performed in 62 outpatients with LVAD (HM 3 N = 31, HW N = 31) and 31 healthy controls. Prevalence and quantity of MES were investigated regarding clinical and laboratory parameters. Cerebrovascular events (CVE) were recorded on follow-up at 90 and 180 days.

RESULTS

MES were detected in six patients with HM 3, three patients with HW, and three controls. Within the LVAD groups, patients on monotherapy with vitamin-K-antagonist (VKA) without antiplatelet therapy were at risk for a higher count of MES (negative binomial regression: VKA: 1; VKA + ASA: Exp(B) = 0.005, 95%CI 0.001-0.044; VKA + clopidogrel: Exp(B) = 0.012, 95%CI 0.002-0.056). There was no association between the presence of MES and CVE or death on follow-up (p > 0.05).

CONCLUSION

For the first time, the prevalence of MES was prospectively investigated in a notable outpatient cohort of patients with HM 3 and HW. Despite optimized properties of the latest LVAD, MES remain detectable depending on antithrombotic therapy. No association between MES and CVE could be detected.

Для цитирования: Ravenberg Kim Kristin, Gabriel Maria Magdalena, Leotescu Andrei, Tran Anh Thu, Grosse Gerrit Maximilian, Schuppner Ramona, Ernst Johanna, Lichtinghagen Ralf, Tiede Andreas, Werwitzke Sonja, Bara Christoph Leon, Schmitto Jan Dieter, Weissenborn Karin, Hanke Jasmin Sarah, Worthmann Hans. Microembolic signal monitoring in patients with HeartMate 3 and HeartWare left ventricular assist devices: Association with antithrombotic treatment and cerebrovascular events.. Artificial organs. 2022. DOI: 10.1111/aor.14409.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.