Development and Internal Validation of an AI-Enabled Cuff-less, Non-invasive Continuous Blood Pressure Monitor Across All Classes of Hypertension.

Lopez-Jimenez Francisco, Deshmukh Abhishek, Bisognano John, Boehmer John, Ramasamy Mouli, Shyam Kumar Prashanth, Kapa Suraj, Varadan Venk, Varadan Vijay, Fudim Marat

СтатьяВалидационное исследованиеМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Non-invasive, continuous blood pressure monitoring technologies require additional validation beyond standard cuff-based methods. This study evaluates a non-invasive, multiparametric wearable cuffless blood pressure (BP) diagnostic monitor across all hypertension classes with diverse subjects.

METHODS

A prospective, multicenter study assessed Nanowear's SimpleSense-BP performance, including induced and natural BP changes, significant BP variations (Systolic BP (SBP) ≥ ± 15 mm Hg and Diastolic BP (DBP) ≥ ± 10 mm Hg), and reference input value validity over 4 weeks.

RESULTS

303 subjects (18-83 yrs; 50.16% Female) participated in algorithmic development and validation (Normal - 35%, Prehypertensive - 24%, Stage 1 - 24%, Stage 2 - 17%). 54 subjects were tested for induced change performance, 149 exhibited significant changes, and 91 validated reference value duration.

CONCLUSIONS

The study clinically validated a continuous, AI-based BP diagnostic monitor using non-invasive wearable data. Further testing on diverse populations and external validation are recommended. The protocol was inspired by ISO 81060-2 and IEEE 1708:2019 standards.

Для цитирования: Lopez-Jimenez Francisco, Deshmukh Abhishek, Bisognano John, Boehmer John, Ramasamy Mouli, Shyam Kumar Prashanth, Kapa Suraj, Varadan Venk, Varadan Vijay, Fudim Marat. Development and Internal Validation of an AI-Enabled Cuff-less, Non-invasive Continuous Blood Pressure Monitor Across All Classes of Hypertension.. Journal of cardiovascular translational research. 2025. DOI: 10.1007/s12265-024-10589-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.