Improved diagnostic sensitivity for tuberculous meningitis with Xpert(®) MTB/RIF of centrifuged CSF.

Bahr N C, Tugume L, Rajasingham R, Kiggundu R, Williams D A, Morawski B, Alland D, Meya D B, Rhein J, Boulware D R

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

BACKGROUND

TB meningitis (TBM) diagnosis is difficult and novel diagnostic methods are needed. The World Health Organization recommends Xpert(®) MTB/RIF as the initial TBM diagnostic test based on two studies reporting suboptimal sensitivity (~50-60%).

OBJECTIVE

To study the effect of cerebrospinal fluid (CSF) centrifugation on Xpert performance for TBM detection.

DESIGN

A total of 107 predominantly human immunodeficiency virus (HIV) infected adults with presumed meningitis were screened prospectively in Kampala, Uganda. CSF was tested using 1) microscopy for acid-fast bacilli; 2) MGIT™ culture; 3) Xpert of 2 ml of unprocessed CSF; and 4) Xpert of centrifuged CSF. Diagnostic performance was measured against an a priori composite reference standard of any positive CSF tuberculosis test.

RESULTS

Of 107 participants, 18 (17%) had definite TBM. When CSF was centrifuged, Xpert had better sensitivity (13/18, 72%) than when using 2 ml of unprocessed CSF (5/18, 28%; P = 0.008). The median centrifuged CSF volume was 6 ml (IQR 4-10). Mycobacterial culture yielded 71% (12/17) sensitivity at a median delay of 27 days. Only 39% were positive by both culture and centrifuged Xpert, with additional cases detected by Xpert and culture.

CONCLUSIONS

CSF centrifugation optimizes the diagnostic performance of Xpert in the detection of TBM. A combination of culture and Xpert detected the largest number of cases.

Для цитирования: Bahr N C, Tugume L, Rajasingham R, Kiggundu R, Williams D A, Morawski B, Alland D, Meya D B, Rhein J, Boulware D R. Improved diagnostic sensitivity for tuberculous meningitis with Xpert(®) MTB/RIF of centrifuged CSF.. The international journal of tuberculosis and lung disease : the official journal of the International Union against Tuberculosis and Lung Disease. 2015. DOI: 10.5588/ijtld.15.0253.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.