Possible Relationship between the Deteriorated Accuracy of Intermittent-Scanning Continuous Glucose Monitoring Device and the Contact Dermatitis: Post-hoc analysis of the ISCHIA Study.

Toyoda Masao, Murata Takashi, Hirota Yushi, Hosoda Kiminori, Kato Ken, Kouyama Kunichi, Kouyama Ryuji, Kuroda Akio, Matoba Yuka, Matsuhisa Munehide, Meguro Shu, Miura Junnosuke, Nishimura Kunihiro, Shimada Akira, Suzuki Shota, Tone Atsuhito, Sakane Naoki, ISCHIA Study Group

Рандомизированное исследованиеМногоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

We previously reported the mean average relative difference (MARD) of the sensor glucose (SG) of the first-generation FreeStyle Libre with the original algorithm, an intermittent scanning continuous glucose monitoring (isCGM) device, was 15.6% in the Effect of Intermittent-Scanning Continuous Glucose Monitoring to Glycemic Control Including Hypoglycemia and Quality of Life of Patients with Type 1 Diabetes Mellitus Study (ISCHIA Study). In the present study, we aimed to further analyze its accuracy in detail by conducting a post-hoc analysis of the study.

METHODS

The ISCHIA Study was a multicenter, randomized, cross-over trial to assess the efficacy of isCGM. The SG levels of isCGM and the measured capillary blood glucose (BG) levels of 91 participants were used for the analysis.

RESULTS

Bland-Altman analysis showed bias of -13.0 mg/dl when the SG levels were compared to the BG levels, however no proportional bias was observed (r = 0.085). MARD of the participants without and with contact dermatitis were 15.0 ± 6.0% and 27.4 ± 21.4% (P = 0.001), respectively.

CONCLUSION

There was negative bias in the SG levels of isCGM compared to the BG levels. There is a possibility that the complication of the contact dermatitis during isCGM use may be related with deteriorated accuracy of the SG levels.

Для цитирования: Toyoda Masao, Murata Takashi, Hirota Yushi, Hosoda Kiminori, Kato Ken, Kouyama Kunichi, Kouyama Ryuji, Kuroda Akio, Matoba Yuka, Matsuhisa Munehide, Meguro Shu, Miura Junnosuke, Nishimura Kunihiro, Shimada Akira, Suzuki Shota, Tone Atsuhito, Sakane Naoki, ISCHIA Study Group. Possible Relationship between the Deteriorated Accuracy of Intermittent-Scanning Continuous Glucose Monitoring Device and the Contact Dermatitis: Post-hoc analysis of the ISCHIA Study.. The Tokai journal of experimental and clinical medicine. 2023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.