Optical quality of foldable monofocal intraocular lenses before and after injection: comparative evaluation using a double-pass system.

Vilaseca Meritxell, Arjona Montserrat, Pujol Jaume, Issolio Luis, Güell José L

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To use the double-pass technique to evaluate the in vitro optical quality of foldable monofocal intraocular lenses (IOLs) used to correct aphakia.

SETTING

Universitat Politècnica de Catalunya, Terrassa, and Instituto de Microcirugía Ocular de Barcelona, Barcelona, Spain; Universidad Nacional de Tucumán, Tucumán, Argentina.

METHODS

This study assessed the in vitro optical quality of 7 IOLs before and after injection in an artificial eye that was attached to a double-pass system (Optical Quality Analysis System [OQAS]). The procedure imitated the conventional in vivo technique used to assess the optical quality of eyes with an IOL. The following parameters were evaluated: point-spread function, modulation transfer function (MTF), MTF cutoff frequency, Strehl ratio, and OQAS values.

RESULTS

The in vitro optical quality of most IOLs was as good after injection as before injection. In 1 IOL, the post-injection optical quality was statistically significantly different but the optical quality remained high.

CONCLUSIONS

Results indicate that after an IOL is placed in the eye, its optical performance will be good, providing good visual quality. The eye cell model attached to the double-pass system was useful and effective for fully characterizing the optical quality of IOLs and evaluating variations resulting from the injection process.

Для цитирования: Vilaseca Meritxell, Arjona Montserrat, Pujol Jaume, Issolio Luis, Güell José L. Optical quality of foldable monofocal intraocular lenses before and after injection: comparative evaluation using a double-pass system.. Journal of cataract and refractive surgery. 2009. DOI: 10.1016/j.jcrs.2009.03.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.