Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality.

He Xiaofang, Nie Chi, Zheng Xiaoyan, Zheng Lei, Wang Chenxiao, Fang Ruqi, Zhong Yumin

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the feasibility of ultra-low tube voltage (70 kV) cardiac CT angiography (CTA) with weight-adapted iodine delivery in neonates and to compare image quality, coronary visibility, radiation exposure, and iodine delivery with those of a conventional 80-kV protocol.

METHODS

This retrospective study included 125 neonates with congenital heart disease who underwent prospectively ECG-triggered cardiac CTA using a 256-row CT scanner. Patients were divided into an 80-kV group reconstructed with 50% adaptive statistical iterative reconstruction-V (ASIR-V) (n = 61) and a 70-kV group reconstructed with 70% ASIR-V (n = 64). Both protocols used standardized weight-adapted iodine delivery (450 mg I/kg). Subjective and objective image quality, coronary artery visibility, radiation dose, and iodine delivery were compared between groups.

RESULTS

Overall interobserver agreement for subjective image quality assessment was excellent (κ = 0.815). Subjective image quality scores were comparable between the two groups across most coronary segments, with no significant differences in overall image quality (3.98 ± 1.04 vs. 3.95 ± 1.05, P = 0.545). Objective image quality parameters, including CT attenuation, image noise, signal-to-noise ratio, and contrast-to-noise ratio, also showed no significant differences between groups (all P > 0.05). Coronary artery visibility remained high in both groups, with overall visibility rates exceeding 89%. Importantly, distal coronary segments remained diagnostically assessable in approximately 75% of cases under the 70-kV protocol. Compared with the 80-kV protocol, the 70-kV protocol achieved a 24% reduction in effective radiation dose (0.32 ± 0.05 vs. 0.42 ± 0.06 mSv, P < 0.001) without increasing iodine delivery.

CONCLUSIONS

Ultra-low tube voltage (70 kV) cardiac CT with weight-adjusted iodine delivery is feasible for neonatal imaging and can substantially reduce radiation exposure while preserving coronary artery image quality and visibility. This approach may help achieve a more balanced trade-off among image quality, radiation dose, and contrast administration in neonatal cardiac CT examinations.

Для цитирования: He Xiaofang, Nie Chi, Zheng Xiaoyan, Zheng Lei, Wang Chenxiao, Fang Ruqi, Zhong Yumin. Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality.. Frontiers in cardiovascular medicine. 2026. DOI: 10.3389/fcvm.2026.1882665.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.