Comparative analysis of platelet counts using Beckman Coulter DxH-900, Mindray BC-6800 Plus, CellaVision DM9600 and the flow cytometry reference method: addressing the challenge of giant platelets.

Piedra-Aguilera Álvaro, Leis-Sestayo Alba, Martínez-Iribarren Alicia, Jiménez-Añón Laura, Rodríguez-Domínguez Jennifer, Cherif-El Kettani Ghali Ech, Tejedor-Ganduxé Xavier, Jurado-Tapiador Rebeca, Morales-Indiano Cristian

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To evaluate the analytical performance of platelet counts obtained by impedance- and optical-based methods on the DxH-900 and BC-6800 Plus analyzers, and by digital morphology using the CellaVision DM9600, in comparison with the international reference method (IRM), in samples with thrombocytopenia and both normal-sized and giant platelets.

METHODS

Platelet counts were analyzed in peripheral blood samples from patients with thrombocytopenia and normal platelet counts, including cases with normal-sized and giant platelets. Each sample was tested using two automated analyzers: the DxH-900 (impedance) and the BC-6800 Plus (impedance and optical). Results were compared with the IRM based on flow cytometry, performed on the DxFlex analyzer. Precision was assessed. The methods were compared by Passing-Bablok regression and bias calculation. Additionally, platelet morphology was reviewed digitally using CellaVision DM9600.

RESULTS

All methods demonstrated high precision and strong correlation with IRM in samples with normal-sized platelets. However, significant discrepancies were observed in the presence of giant platelets. Impedance methods showed higher imprecision (CV>10 %) and underestimated platelet counts with a negative bias exceeding -25 %. Conversely, the optical method showed better correlation (r=0.9888) and precision (CV=1.6 %), although it overestimated counts with a positive bias of 19.4 %. Digital morphology also demonstrated strong agreement with IRM.

CONCLUSIONS

These findings highlight the analytical limitations of routine technologies in challenging samples and emphasize the importance of method selection for accurate platelet reporting. A correction factor for optical counts and the integration of digital morphology could enhance diagnostic reliability in cases involving giant platelets.

Для цитирования: Piedra-Aguilera Álvaro, Leis-Sestayo Alba, Martínez-Iribarren Alicia, Jiménez-Añón Laura, Rodríguez-Domínguez Jennifer, Cherif-El Kettani Ghali Ech, Tejedor-Ganduxé Xavier, Jurado-Tapiador Rebeca, Morales-Indiano Cristian. Comparative analysis of platelet counts using Beckman Coulter DxH-900, Mindray BC-6800 Plus, CellaVision DM9600 and the flow cytometry reference method: addressing the challenge of giant platelets.. Advances in laboratory medicine. 2025. DOI: 10.1515/almed-2025-0133.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.