Baumann S, Becher T, Giannakopoulos K, Jabbour C, Fastner C, El-Battrawy I, Ansari U, Lossnitzer D, Behnes M, Alonso A, Kirschning T, Dissmann R, Kueck O, Stern D, Michels G, Borggrefe M, Akin I

Аннотация

BACKGROUND: Pulmonary embolism (PE) is a frequently occurring complication in critically ill patients, and the simultaneous occurrence of PE and life-threatening bleeding is a therapeutic dilemma. Inferior vena cava filters (IVCF) may represent an important therapeutic alternative in these cases. The Angel® catheter (Bio2 Medical Inc., San Antonio, TX, USA) is a novel IVCF that provides temporary protection from PE and is implanted at bedside without fluoroscopy.

MATERIAL AND METHODS: The European Angel® Catheter Registry is an observational, multicenter study. In our German substudy, we investigated patients from three German hospitals and four intensive care units, who underwent Angel® catheter implantation between February 2016 and December 2016.

RESULTS: A total of 23 critically ill patients (68 ± 9 years, 43% male) were included. The main indication for implantation was a high risk for or an established PE, combined with contraindications for prophylactic or therapeutic anticoagulation due to either an increased risk of bleeding (81%) or active bleeding (13%). The Angel® catheter was successfully inserted in all patients at bedside. No PE occurred in patients with an indwelling Angel® catheter. Clots with a diameter larger the 20 mm, indicating clot migration, were detected in 5% of the patients by cavography before filter retrieval. Filter retrieval was uneventful in all of our cases, while filter dislocation occurred in 3% of the patients.

CONCLUSION: The German data from the multicenter European Angel® Catheter Registry show that the Angel® catheter is a safe and effective approach for critically ill patients with a high risk for the development of PE or an established PE, when an anticoagulation therapy is contraindicated.