DOI 10.1097/ijg.0000000000000784
Reversible Venting Stitch for Fenestrating Valve-less Glaucoma Shunts.
Аннотация
PURPOSE: The purpose of this is to describe a venting stitch modification for valveless glaucoma aqueous shunts and characterize early postoperative intraocular pressure (IOP) and glaucoma medication use following the modification.
MATERIALS AND
METHODS
Retrospective chart review of 61 sequential patients undergoing Baerveldt glaucoma implant (BGI)-350 implantation at the Doheny Eye Institute. Twenty-four patients received a glaucoma shunt with venting stitch modification (modified BGI) and 37 patients received an unmodified shunt (BGI-only). IOP, number of glaucoma medications, and number of hypotony cases (intraocular pressure ≤5 mm Hg) were compared between the groups. T-tests were used for statistical analysis.
RESULTS
At postoperative-day 1, mean IOP was significantly lower compared with preoperatively in the modified BGI group (14 mm Hg; reduced by 51%; P<0.0001) but not the BGI-only group (27 mm Hg; P=0.06). IOP difference between groups persisted till immediately before tube opening (P=0.005) and fewer IOP-lowering medications needed in the modified BGI group (P<0.0001). One case (4.2%) of postoperative hypotony was encountered with BGI modification, which resolved after the stitch was removed in clinic.
CONCLUSIONS
The venting stitch valveless shunt modification allows for effective, reliable, and safe control of early postoperative IOP.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.