MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jvim.16917

Медицинская библиотека · Journal of veterinary internal medicine · 2023

Accuracy of Crit-Line® monitors in predicting hematocrit and change in blood volume of dogs during hemodialysis.

Vickery Whitney E, Steinbach Sarah M L, Moore George E

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Ultrafiltration (UF) is an extracorporeal technique for treating fluid overload and is monitored with noninvasive blood volume monitors.

HYPOTHESIS/

OBJECTIVES

The primary objective was to determine the accuracy of Crit-Line® III and IV noninvasive blood volume monitors to estimate canine packed cell volume (PCV) and changes in blood volume during UF. A secondary aim was to determine accuracy of targeted ultrafiltration rates (UFR) compared to calculated delivered UFR.

METHODS

An ex vivo study with a single Phoenix® X36 platform and canine packed red blood cells (pRBC) was performed. Through dilution and UF, clinically applicable PCV values were obtained and compared to Crit-Line® hematocrit (Hct). Blood volume was constantly measured and compared to targeted UFR. Systematic and proportional bias were calculated using a Bayes method.

RESULTS

Crit-Line® III and IV reported Hct was significantly lower than PCV (n = 140, median 26%, range, 8%-50%) when PCV was >25% and >30%, respectively. Crit-Line® III and IV calculated change in blood volume (ΔBV%) was significantly different from measured blood volume changes at ΔBV% ±20% and >-20%, respectively. Comparing targeted and delivered UFR (mL/h), less than targeted UFR was removed at UFR100 through UFR400 and UFR0 adding volume.

CONCLUSIONS AND CLINICAL

IMPORTANCE

Crit-Line® III and IV monitors provide accurate estimates of canine PCV and UF volume change within specific ranges and are useful in monitoring canine UF and hemodialysis. Veterinary hemodialysis services should evaluate individual machines for UFR inaccuracies, which can meaningfully affect small animals.

Для цитирования: Vickery Whitney E, Steinbach Sarah M L, Moore George E. Accuracy of Crit-Line® monitors in predicting hematocrit and change in blood volume of dogs during hemodialysis.. Journal of veterinary internal medicine. 2023. DOI: 10.1111/jvim.16917.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.