DOI 10.1016/j.jvir.2018.01.002
Evaluation of a CT-Guided Robotic System for Precise Percutaneous Needle Insertion.
Аннотация
PURPOSE: To assess overall targeting accuracy for CT-guided needle insertion using prototype robotic system for common target sites.
MATERIALS AND
METHODS
Using CT guidance, metallic (2 × 1 mm) targets were embedded in retroperitoneum (n = 8), kidneys (n = 8), and liver (n = 14) of 8 Yorkshire pigs (55-65 kg). Bronchial bifurcations were targeted in the lung (n = 13). CT datasets were obtained for planning and controlled needle placement of commercially available 17- to 19-gauge needles (length 15-20 cm) using a small, patient-mounted, CT-guided robotic system with 5° of motion. Mean distance to target was 92.9 mm ± 19.7 (range, 64-146 mm). Planning included selection of target, skin entry point, and 4.6 ± 1.3 predetermined checkpoints (range, 2-9) where additional CT imaging was performed to permit stepwise correction of needle trajectory path as needed. Scanning and needle advancement were coordinated with breath motion using respiratory gating. Accuracy was assessed as distance from needle tip to predefined target.
RESULTS
Of 45 needle insertions performed, 2 were unsuccessful owing to technical issues. Accuracy of targeting was 1.2-1.4 mm ± 0.6 for kidney, retroperitoneum, and lung (P = .51), with 2.9 mm ± 1.9 accuracy for liver (P = .0003). This was achieved in 39 cases (91%) using a single insertion. Intraprocedural target movement was detected (3.5 mm ± 2.1 in retroperitoneum and 6.4 mm ± 3.9 in liver); the system compensated for 52.9% ± 30.3 of this movement. One pneumothorax was the only complication (8%).
CONCLUSIONS
Accurate needle insertion (< 3 mm error) can be achieved in common target sites when using a CT-guided robotic system. Stepwise checks with corrective angulation can potentially overcome issues of target movement during a procedure from organ deformity and other causes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.