Investigative radiology · 2019
Breast Cancer Assessment With Pulse-Echo Speed of Sound Ultrasound From Intrinsic Tissue Reflections: Proof-of-Concept.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ruby Lisa, Sanabria Sergio J, Martini Katharina, Dedes Konstantin J, Vorburger Denise, Oezkan Ece, Frauenfelder Thomas, Goksel Orcun, Rominger Marga B
Аннотация
PURPOSE: The aim of this study was to differentiate malignant and benign solid breast lesions with a novel ultrasound (US) technique, which measures speed of sound (SoS) using standard US transducers and intrinsic tissue reflections and scattering (speckles) as internal reference.
MATERIALS AND METHODS: This prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant prospective comparison study was performed with prior written informed consent from 20 women. Ten women with histological proven breast cancer and 10 with fibroadenoma were measured. A conventional US system with a linear probe was used for SoS-US (SonixTouch; Ultrasonix, Richmond, British Columbia, Canada). Tissue speckle reflections served as a timing reference for the US signals transmitted through the breasts. Relative phase inconsistencies were detected using plane wave measurements from different angular directions, and SoS images with 0.5-mm resolution were generated using a spatial domain reconstruction algorithm. The SoS of tumors were compared with the breast density of a larger cohort of 106 healthy women.
RESULTS: Breast lesions show focal increments ΔSoS (meters per second) with respect to the tissue background. Peak ΔSoS values were evaluated. Breast carcinoma showed significantly higher ΔSoS than fibroadenomas ([INCREMENT]SoS > 41.64 m/s: sensitivity, 90%; specificity, 80%; area under curve, 0.910) and healthy breast tissue of different densities (area under curve, 0.938; sensitivity, 90%; specificity, 96.5%). The lesion localization in SoS-US images was consistent with B-mode imaging and repeated SoS-US measurements were reproducible.
CONCLUSIONS: Using SoS-US, based on conventional US and tissue speckles as timing reference, breast carcinoma showed significantly higher SoS values than fibroadenoma and healthy breast tissue of different densities. The SoS presents a promising technique for differentiating solid breast lesions.
MATERIALS AND METHODS: This prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant prospective comparison study was performed with prior written informed consent from 20 women. Ten women with histological proven breast cancer and 10 with fibroadenoma were measured. A conventional US system with a linear probe was used for SoS-US (SonixTouch; Ultrasonix, Richmond, British Columbia, Canada). Tissue speckle reflections served as a timing reference for the US signals transmitted through the breasts. Relative phase inconsistencies were detected using plane wave measurements from different angular directions, and SoS images with 0.5-mm resolution were generated using a spatial domain reconstruction algorithm. The SoS of tumors were compared with the breast density of a larger cohort of 106 healthy women.
RESULTS: Breast lesions show focal increments ΔSoS (meters per second) with respect to the tissue background. Peak ΔSoS values were evaluated. Breast carcinoma showed significantly higher ΔSoS than fibroadenomas ([INCREMENT]SoS > 41.64 m/s: sensitivity, 90%; specificity, 80%; area under curve, 0.910) and healthy breast tissue of different densities (area under curve, 0.938; sensitivity, 90%; specificity, 96.5%). The lesion localization in SoS-US images was consistent with B-mode imaging and repeated SoS-US measurements were reproducible.
CONCLUSIONS: Using SoS-US, based on conventional US and tissue speckles as timing reference, breast carcinoma showed significantly higher SoS values than fibroadenoma and healthy breast tissue of different densities. The SoS presents a promising technique for differentiating solid breast lesions.