Marais Ben J, Brittle Wendy, Painczyk Katrien, Hesseling Anneke C, Beyers Nulda, Wasserman Elizabeth, van Soolingen Dick, Warren Rob M
Аннотация
BACKGROUND: Fluorescence microscopy offers well-described benefits, compared with conventional light microscopy, for the evaluation of sputum smear samples for tuberculosis. However, its use in resource-limited settings has been limited by the high cost of the excitatory light source. We evaluated the diagnostic performance of fluorescence microscopy, using novel light-emitting diode (LED) technology as an alternative to the conventional mercury vapor lamp (MVP).
METHODS: Routinely collected sputum specimens from persons suspected to have tuberculosis who attended community clinics were stained with auramine O and were evaluated using 2 different excitatory light sources (MVP and LED); these specimens were then Ziehl-Neelsen stained and reexamined using light microscopy. Two microscopists independently evaluated all smears. Bacterial culture provided the gold standard.
RESULTS: Of the 221 sputum specimens evaluated, 36 (16.3%) were positive for Mycobacterium tuberculosis by culture. Sensitivity and specificity documented for the different modalities were 84.7% and 98.9%, respectively, for the LED assessment; 73.6% and 99.8%, respectively, for the MVP assessment; and 61.1% and 98.9%, respectively, for light microscopy. kappa values for interreader variation were 0.87 for the LED assessment, 0.79 for the MVP assessment, and 0.77 for light microscopy. The mean time to read a negative smear was 1.4 min with fluorescence microscopy and 3.6 min with light microscopy, reflecting a time savings of 61% with fluorescence microscopy.
CONCLUSION: LED fluorescence microscopy provides a reliable alternative to conventional methods and has many favorable attributes that facilitate improved, decentralized, diagnostic services.
METHODS: Routinely collected sputum specimens from persons suspected to have tuberculosis who attended community clinics were stained with auramine O and were evaluated using 2 different excitatory light sources (MVP and LED); these specimens were then Ziehl-Neelsen stained and reexamined using light microscopy. Two microscopists independently evaluated all smears. Bacterial culture provided the gold standard.
RESULTS: Of the 221 sputum specimens evaluated, 36 (16.3%) were positive for Mycobacterium tuberculosis by culture. Sensitivity and specificity documented for the different modalities were 84.7% and 98.9%, respectively, for the LED assessment; 73.6% and 99.8%, respectively, for the MVP assessment; and 61.1% and 98.9%, respectively, for light microscopy. kappa values for interreader variation were 0.87 for the LED assessment, 0.79 for the MVP assessment, and 0.77 for light microscopy. The mean time to read a negative smear was 1.4 min with fluorescence microscopy and 3.6 min with light microscopy, reflecting a time savings of 61% with fluorescence microscopy.
CONCLUSION: LED fluorescence microscopy provides a reliable alternative to conventional methods and has many favorable attributes that facilitate improved, decentralized, diagnostic services.