Use of light-emitting diode fluorescence microscopy to detect acid-fast bacilli in sputum.
Аннотация
BACKGROUND
Fluorescence microscopy offers well-described benefits, compared with conventional light microscopy, for the evaluation of sputum smear samples for tuberculosis. However, its use in resource-limited settings has been limited by the high cost of the excitatory light source. We evaluated the diagnostic performance of fluorescence microscopy, using novel light-emitting diode (LED) technology as an alternative to the conventional mercury vapor lamp (MVP).
METHODS
Routinely collected sputum specimens from persons suspected to have tuberculosis who attended community clinics were stained with auramine O and were evaluated using 2 different excitatory light sources (MVP and LED); these specimens were then Ziehl-Neelsen stained and reexamined using light microscopy. Two microscopists independently evaluated all smears. Bacterial culture provided the gold standard.
RESULTS
Of the 221 sputum specimens evaluated, 36 (16.3%) were positive for Mycobacterium tuberculosis by culture. Sensitivity and specificity documented for the different modalities were 84.7% and 98.9%, respectively, for the LED assessment; 73.6% and 99.8%, respectively, for the MVP assessment; and 61.1% and 98.9%, respectively, for light microscopy. kappa values for interreader variation were 0.87 for the LED assessment, 0.79 for the MVP assessment, and 0.77 for light microscopy. The mean time to read a negative smear was 1.4 min with fluorescence microscopy and 3.6 min with light microscopy, reflecting a time savings of 61% with fluorescence microscopy.
CONCLUSION
LED fluorescence microscopy provides a reliable alternative to conventional methods and has many favorable attributes that facilitate improved, decentralized, diagnostic services.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.