m+Medelevator.RU
← Поиск публикаций
Abdominal radiology (New York) · 2024

Achieving sub-millisievert CT colonography for accurate colorectal tumor detection using smart examination protocols: a prospective self-controlled study.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zhang Jingyi, Hu Mengting, Cheng Qiye, Wang Shigeng, Liu Yijun, Zhou Yujing, Li Jianying, Wei Wei

Аннотация

PURPOSE: To assess the feasibility of combining Auto-kVp selection technique, higher preset ASIR-V and noise index (NI) to realize individualized sub-mSv CT colonography (CTC) for accurate colorectal tumor detection and localization.

METHODS: Ninety patients with suspected colorectal cancer (CRC) were prospectively enrolled to undergo standard dose CTC (SDCTC) in the prone and ultra-low dose CTC (ULDCTC) in the supine position. SDCTC used 120 kVp, preset ASIR-V of 30%, SmartmA for a NI of 13; ULDCTC used Auto-kVp selection technique with 80 or 100 kVp, preset ASIR-V of 60%, SmartmA for a NI of 13 for 80 kVp, and NI of 15 for 100 kVp. The effective dose (ED), image quality [signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of colorectal neoplasms] between the two protocols were compared and the accuracies of tumor locations were evaluated for CTC in comparison with the surgery results.

RESULTS: The mean ED of the ULDCTC-80 kVp subgroup was 0.70 mSv, 71.43% lower than the 2.45 mSv for the 120 kVp group, while that of the ULDCTC-100 kVp subgroup was 0.98 mSv, 73.00% lower than the 3.63 mSv for the 120 kVp group (P < 0.001). The tumor SNR and CNR of the ULDCTC were higher than those of SDCTC (P < 0.05), while there was no difference in the subjective image quality between them with good inter-observer agreement (Kappa: 0.805-0.923). Both SDCTC and ULDCTC groups had high detection rate of colorectal tumors, along with good consistency in determining tumor location compared with surgery reports (Kappa: 0.718-0.989).

CONCLUSION: The combination of Auto-kVp selection, higher preset ASIR-V and NI achieves individualized sub-mSv CTC with good performance in detecting and locating CRC with surgery and consistent results between SDCTC and ULDCTC.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты