MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/cej.0000000000000887

Медицинская библиотека · European journal of cancer prevention : the official journal of the European Cancer Prevention Organisation (ECP) · 2024

Non-contrast MRI and post-mastectomy silicone breast implant rupture: preventing false positive diagnoses.

Mazzocconi Luca, De Lorenzi Francesca, Carbonaro Riccardo, Lorenzano Valerio, Rotili Anna, Pesapane Filippo, Signorelli Giulia, Caldarella Pietro, Corso Giovanni, Cassano Enrico, Veronesi Paolo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Breast implants are not lifelong, with implant rupture being the third leading cause of revisional surgery in augmented women. Noncontrast MRI is a reliable tool to assess implant integrity; however, false positive and false negative diagnoses have been reported due to an incorrect interpretation of MRI signs. This study aims to investigate the incidence of these misleading results, comparing MRI findings with intraoperative surgical observations and exploring signs of nonunivocal interpretation.

MATERIALS AND

METHODS

Between March 2019 and October 2022, our hospital, a referral center for breast cancer care, conducted 139 breast MRI examinations to evaluate implant integrity. Surgical intervention was deemed necessary for patients diagnosed with suspected or confirmed implant rupture at MRI. Those patients who did not undergo any surgical procedure (63 cases) or had surgery at different institutes (11 cases) were excluded.

RESULTS

Among the 65 patients who underwent preoperative MRI and subsequent surgery at our institute, surgical findings confirmed the preoperative MRI diagnosis in 48 women. Notably, 17 women exhibited a discordance between MRI and surgical findings: three false negatives, 11 false positives and three possible ruptures not confirmed. Signs of nonunivocal or misleading interpretation were assessed on a patient-by-patient basis. The importance of obtaining detailed information about a patient's breast implant, including fill materials, number of lumens, manufacturer and shape, proved immensely beneficial for interpreting MRI signs accurately.

CONCLUSION

Pre-MRI knowledge of implant details and a meticulous evaluation of non-univocal signs can aid radiologists in accurately assessing implant integrity, reducing the risk of unnecessary revisional surgeries, and potentially averting allegations of medical malpractice.

Для цитирования: Mazzocconi Luca, De Lorenzi Francesca, Carbonaro Riccardo, Lorenzano Valerio, Rotili Anna, Pesapane Filippo, Signorelli Giulia, Caldarella Pietro, Corso Giovanni, Cassano Enrico, Veronesi Paolo. Non-contrast MRI and post-mastectomy silicone breast implant rupture: preventing false positive diagnoses.. European journal of cancer prevention : the official journal of the European Cancer Prevention Organisation (ECP). 2024. DOI: 10.1097/cej.0000000000000887.